TW-150

Water Distribution Monitor

HORIBA's TW-150 is a process water quality measurement system designed for drinking water and water distribution plants. TW-150 monitors turbidity,color, free residual chlorine, water pressure and pH without any reagent for measurement.
This chemical free method reduces operating costs as it requires minimal maintenance and improves process control.

Division: Water Pollution
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Turbidity Method

Measurement method for turbidity is direct NTU value by 90 degree light scattering method and comply with ISO 7027.
Measurement of turbidity and color is performed with a single cell (patent pending).

Standard

5 parameters
Turbidity, Color, Free residual chlorine,
Water pressure, pH

Option

7 parameters
Turbidity, Color, Free residual chlorine,
Water pressure, pH, Conductivity,
Water temperature

Touch panel operation

The instrument is operated with a touch screen panel for easy operation and display of measured values.

●Operation panel specifications
320 x 240 dots
Black and white graphic LCD with backlight

 

User-friendly function

●The adoption of transparent measuring cell enables sample conditions to be easily confirmed.

 

 

●The use of thumbscrews makes maintenance easier as special tools are not required.

 

Product name

Water Distribution Monitor

Type

TW-150

Measurement items(Standard configuration) turbidity, color, free residual chlorine, water pressure, pH
(Optional items) conductivity, water temperature
Measurement Method

See Table 1.

Measurement  Range

See Table  1.

Repeatability

See Table  1.

Display system

LCD display: 320 x 240 dot backlit change to five graphic liquid-crystal display (touch panel type)
Up to seven items including five items from standard configuration (turbidity, color, free residual chlorine, water pressure, pH)
and optional items (conductivity, water temperature) can be displayed at the same time.

Calibration method

See Table 1.

Automatic zero calibration

(Turbidity, color, free residual chlorine)
Calibration method: Filtrate sample water
Calibration start: Interior…started by internal timer
Exterior…started by external contact input
Calibration cycle: 5 hours to 9999 hours (user setting)
Calibration time: Approx. 15 minutes

Automatic cleaning

(Turbidity, color)
Cleaning method: Cell window cleaning using wiper
Cleaning start: Interior...started by internal timer
Exterior...started by external contact input
Cleaning interval: 5 to 9999 minutes (user setting)
(Free residual chlorine)
Continuous cleaning by using grinding beads

Automatic drainingDraining start: Interior...started by internal timer
Draining interval: 5 to 9999 minutes (user setting)

Sample water conditions

Temperature: 0°C to 40°C (do not freeze)
Pressure: 0.1 MPa to 0.75 MPa
Conductivity: 10 mS/m or higher
Analyzing unit input rate (flow rate): 50 mL/mi to 100 mL/min
● In test operation, flush the meter thoroughly before running water.
● Make sure to set up a bypass for piping to the meter.
● If sample water may freeze, take measures to insulate the unit from cold and to retain heat.
● Sample Water measured with this device cannot be distributed.

Ambient temperature,humidity

Ambient temperature: 0°C to 40°C
Ambient humidity: 85% or lower

Analog output

Type: Measurement values: Number of outputs corresponds to number of measurement values (up to seven values including
five values of the standard configuration)
Specifications: 4 mA to 20 mA DC, insulated output (non-insulated between items)
Maximum load resistance: 600 Ω
Contact output Types: Power failure, batch alarm, maintenance
Contents: Power failure…occurs when the power is turned off
Batch alarm…cell temperature adjustment abnormality, temperature compensation abnormality,
cell wiper abnormality internal communication abnormality, leak, battery abnormality,
concentration upper limit concentration lower limit,concentration device upper limit,
concentration device lower limit, light source abnormality zero calibration, span calibration, analyzer abnormality
Maintenance…when the system enters maintenance or calibration mode
Specifications: No-voltage contact output, a contact interface
Contact rating: 125 V AC 0.3 A, 30 V DC 1 A (at resistance load)
Each output is an independent COM interface.

Contact input

Type: Cleaning start, zero calibration start, alarm check setting, abnormal water sampling
Contents: Cleaning start…started by closed contact input
Zero calibration start…started by closed contact input
Alarm check setting…started by closed contact input
Abnormal water sampling... started by closed contact input
Specifications: No-voltage contact input (open collector connection is possible), insulated input
ON resistance: maximum 100 Ω
Open voltage: 24 V DC
Short-circuit current: maximum 13 mA

Communications

Interface: RS-232C compatible
Communication speed: 19200 bps

Functions

Integrating function for flow rate used in the system (counting type)
Internal leak detecting function

Data memory

Measurement data is stored on system, and can be transferred to a CompactFlash®* (CF Card).
Data saving interval: 1 minute or 1 hour
Data saving frequency: 1 minute: Saves at every 0 second of the minute
1 hour: Saves at every 0 second of the hour
Data memory time: 1 minute interval…for approx. 10 days
1 hour interval…for approx. 1 year
The latest data will be stored.
Wiring connectorWiring inlet
Compliant cable: 12.5 mm dia. to 14.5 mm dia.
Power source: 4.5 mm to 6 mm
Pipe connectorSample inlet: Rc1/4
Condensate outlet: Rc1/8
Drain: Rc1/4
Condensate outlet (internal): 6 mm dia. hose nipple
Air inlet: Rc1/8
Condensate outlet (for detection): 6 mm dia. hose nipple (Rc1/8)
Calibration inlet: Rc1/8
InstallationDesigned for indoor installation.
* For outdoor installation, a case is required (option).
Power supply90 V to 264 V AC: 50/60 Hz

Power consumption

100 V to 240 V AC: Maximum 120 VA

Weight

Approx. 18 kg 

Dimensions

350 (W) x 160 (D) x 420 (H) (unit: mm)

Paint color

Munsell 5PB 8/1

Installation environment

● Flat and stable location with minimum vibrations or shocks
● No dust, mist or corrosive gas in the air
● Under atmospheric pressure
● No direct sunlight
● Good ventilation
● Altitude 2000 m or lower

* CompactFlash® is a registered trademark or trademark of SanDisc Corporation in the United States and other countries.


[Table 1]
■Standard type

Measurement itemsMeasurement methodMeasurement rangeRepeatabilityCalibration method
Turbidity90 degree light scattering method0 to 2,0 to 5, 0 to 10 NTU+2.5% of full scaleFormazin standard solution
ColorTransmitted light absorption method0 to 10,0 to 20 TCU+5.0% of full scaleStandard color solution
Free residual chlorinePolarographic method0 to 5 mg/L+2.5% of full scaleDPD colorimetric method
Water pressureSemiconductor detection method0 to 1 MPa, 0 to 10 Bar+1.0% of full scaleStandard pressure gauge
pHGlass electrode method2 to 12 pH+0.1 pHpH standard solution


■Option (each parameter can be added to the standard type)

Measurement itemsMeasurement methodMeasurement RangeRepeatabilityCalibration method
ConductivityAC 2 pole method0 to 500, 0 to 1000 μS/cm+2.0% of full scaleKCI standard solution
Water temperatureThermistor methodO to 50°C+0.5°CStandard thermometer

Request for Information

Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.

Related products

Aqualog
More Aqualog

Water Treatment Plant Analyzer

Aqualog Datastream
More Aqualog Datastream

For Instant Water Quality Reports

CS-600F
More CS-600F

Fiber Optic Type Chemical Concentration Monitor

CS-620F
More CS-620F

Optical Fiber Type Hot Phosphoric Acid Concentration Monitor

CS-900
More CS-900

Non-Contact Chemical Concentration Monitor

ESH-01-C-SN
More ESH-01-C-SN

Sanitary Conductivity Sensor (Ferrule Clamp type) for Low Concentrations

FES-310L
More FES-310L

Sanitary Conductivity Sensor (Flow Through type)

FS-01FC-SL Series
More FS-01FC-SL Series

Sanitary Conductivity Sensor (Flow Through type) for Low Concentrations

Glass Electrode 1076A-10C
More Glass Electrode 1076A-10C

For measurement of low-conductivity water and non-aqueous solvents.

HC-200F (Four-Wire Analyzer)
More HC-200F (Four-Wire Analyzer)

Industrial simplified fluoride ion concentration meter

HC-200NH
More HC-200NH

Ammonia Nitrogen Meter

HC-300F (Two-Wire Transmitter)
More HC-300F (Two-Wire Transmitter)

Industrial simplified fluoride ion concentration meter

HD-200
More HD-200

Industrial dissolved oxygen analyzer (Four-Wire Analyzer)

HD-200FL
More HD-200FL

Industrial dissolved oxygen analyzer (Four-Wire Analyzer)

HD-300
More HD-300

Industrial dissolved oxygen analyzer (Two-Wire Transmitter)

HD-480
More HD-480

Industrial dissolved oxygen analyzer (The Slim48 series)

TW-100
More TW-100

Water Distribution Monitor