Bromide ion electrode 8005-10C

Bromide ion electrode 8005-10C

 

 

Segment: Scientific
Division: Water Quality
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.
Measuring range0.8 to 80,000 mg/L Br-
(10-5 to 1 mol/L) Br-
pH range800 mg/L (10-2 mol/L) Br-
pH 1.5 to 11.5
Applicable temperature range0 to 50℃
Response time (90%)Within 5 seconds
Applicable reference electrode2565A
Selection coefficientS2O32-、l-、S2-、CN- = Not acceptable
MnO4- = 1
Cl-、PO42- = 100
F-、NO3-、SO42- = 1,000

• The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.
Exception:
*1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L

• The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.

Related products

Aqualog - Water Treatment Plant Analyzer
Aqualog - Water Treatment Plant Analyzer

Automated Organic Analysis & Early Warning Sentinel

Aqualog Datastream
Aqualog Datastream

For Instant Water Quality Reports

LAQUA F-72
LAQUA F-72

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-73
LAQUA F-73

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-74
LAQUA F-74

Benchtop pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

514-05
514-05

0.5% NaCl Standard Solution

514-22
514-22

1.41 mS/cm Conductivity Standard Solution

514-23
514-23

12.9 mS/cm Conductivity Standard Solution

514-50
514-50

5.0% NaCl Standard Solution