HD-300

Field-installation type dissolved oxygen meter (DO meter)

For control and management in wastewater treatment and production processes.

Ideal for the management of aeration tanks at facilities such as wastewater treatment plants and industrial wastewater treatment facilities. Featuring a low-cost membrane cap replacement system and safe inner liquids with KCl (potassium chloride).

DO

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Low-cost sensor replacement

For sensor replacement, only the membrane cap needs to be replaced, making it more cost-effective compared to replacing the sensor chip.

Safe inner liquid

Utilizing a safe KCl solution as the inner liquids.

Product name Dissolved oxygen meter
Model HD-300
Combined sensor 5505, 5510 (A bipolar diaphragm type polarography sensor)
Combined probe DO-1100
Measurement
range
Dissolved oxygen
concentration
0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L)
Saturation degree 0% to 200% (Display range: 0% to 200%)
Temperature 0℃ to 50℃ (Display range: -10℃ to 110℃)
Display
resolution
Dissolved oxygen
concentration
0.01 mg/L
Saturation degree 1%
Temperature 0.1℃
Performance Dissolved
oxygen
concentration
Repeatability Within ±0.5% of the full scale (response for equivalent input)
Linearity Within ±0.5% of the full scale (response for equivalent input)
Temperature Repeatability ±0.5℃ (response for equivalent input)
Linearity ±0.5℃ (response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 600 Ω(when power supply voltage is 24 V DC) (*1)
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact input Number of input points 1
Contact type No-voltage a-contact
Conditions ON resistance: 40 Ω maximum.
Open voltage: 1.2 V DC
Short-circuit current: 21 mA DC
Contact function External input for transmission holding.
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
(The temperature sensor is built into the probe DO-1000.)
Temperature compensation
range
0℃ to 50℃
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Calibration method Span calibration in the air or DO saturated aquaous solution
Number of calibration points Atmospheric calibration: 1 point
(Zero electric calibration is carried out automatically)
Saturation liquid calibration: 2 points (zero calibration is omitted)
Additional capabilities Salinity compensation (0.0% to 5.0%)
Automatic detection of calibration failure (Zero error, Sensitivity error)
Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity)
Self-check Calibration error Zero calibration error, sensitivity error, outside of temperature calibration range
Electrode diagnostic error Sensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor
meter error CPU error, CPU error, ADC error, memory error
Operating temperature range -20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature -25℃ to 65℃
Power supply Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC)*1
Power consumption 0.6 W max.
Compatible
standards
CE marking EMC  :EN61326-1
     Class A, Industrial electromagnetic environment
RoHS  :EN50581
     9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Attached to 50A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 (Epoxy degenerated melamine resin painting)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 2.8 kg
Cover and mounting brackets: Approx. 1 kg

*1: The maximum load resistance that can be connected is decided depending on the power-supply voltage.

Relation between power supply voltage and load resistance

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.