C/S analyzers refers to products that measure Carbon and Sulfur in inorganic materials. Alternative denomination is Elemental Analyzers even if only 2 elements are measured. ISO standards talk about Infrared Absorption Method after combustion. HORIBA uses the acronym EMIA, which stands for Element Material Infrared Analyzer.
O/N/H analyzers refers to products that measure Oxygen, Nitrogen, and Hydrogen in inorganic materials. Depending on the model, 1, 2 or 3 gases are measured. Alternative denomination is also Elemental Analyzers. ISO standards describe the principle as Inert Gas Fusion Technique. HORIBA uses the acronym EMGA, which stands for Element Material Gas Analyzer.
C/S & O/N/H analyzers measure inorganic solid samples such as small blocks, chips, powders or granulates. The principle of these instruments is to heat the sample to very high temperature to ensure complete combustion and to carry the gases generated to detectors where the elements will be measured. Therefore, EMIA & EMGA combine high performance furnaces and specific detection systems.
For EMIA, a flow of Oxygen is used to carry CO, CO2 and SO2 to the detection system while for EMGA a flow of inert gas is used to carry H2, N2 and CO (produced with reaction of oxygen with a graphite crucible).
There are 3 different types of furnaces depending on the instrument and the application:
The table below summarizes the principle for each element and instrument.
Element | Carrier gas | Furnace | Crucible | Principle | Detection technique | Model |
---|---|---|---|---|---|---|
C | O2 | Induction | Ceramic crucible | Oxidation / Generates CO and CO2 | NDIR | EMIA Pro / Expert |
Resistance | Ceramic boat | EMIA Step | ||||
S | O2 | Induction | Ceramic crucible | Oxidation / Generates SO2 | NDIR | EMIA Pro / Expert |
Resistance | Ceramic boat | EMIA Step | ||||
O | He | Impulse | Graphite crucible | Inert gas fusion, reduction / Generates CO and CO2 | NDIR | EMGA Pro / Expert EMGA 930 / 920 |
N | He | Graphite crucible | Inert gas fusion, reduction / Generates N2 | TCD | EMGA Pro / Expert EMGA 930 / 920 | |
H | He | Graphite crucible | Inert gas fusion, reduction / Generates H2 + conversion to H2O | NDIR | EMGA Pro / Expert EMGA 930 | |
Ar | Graphite crucible | Inert gas fusion, reduction / Generates H2 | TCD | EMGA Pro / Expert EMGA 921 |
Below are synopsis of EMIA and EMGA instruments
EMIA C/S analyzer synopsis
EMGA O/N/H analyzer synopsis
EMGA H analyzer synopsis
Do you have any questions or requests? Use this form to contact our specialists.
Carbon/Sulfur Analyzer
Carbon/Sulfur Analyzer
Carbon/Sulfur Analyzer
Oxygen/Nitrogen/Hydrogen Analyzers (Entry Model)
Oxygen/Nitrogen/Hydrogen Analyzers
(Flagship High-Accuracy Model)
Oxygen/Nitrogen Analyzer
Hydrogen Analyzer
Oxygen/Nitrogen/Hydrogen Analyzer