EMGA-920

Overview

EMGA-920 O/N analyzer: It's a "New flagship model” for O/N analyzers and user friendly product.

This is an oxygen and nitrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on.   This is a new generation model optimized to fit to user's requests.

The oxygen is measured as carbon monoxide and carbon dioxide by a non-dispersive infrared detector and the nitrogen by a thermal conductivity detector.


Manufactured by HORIBA

Specifications

EMGA-920:Oxygen/Nitrogen analyzer

Principle
Oxygen:Non Dispersive Infrared detector
Nitrogen:Thermal Conductivity detector(TCD)

Measurement range*:
Oxygen:0-5%, Nitrogen:0-3%
*Up to 100% wt is possible by decreasing the sample weight.

Sample weight: 1,0 ± 0,1g

Sensitivity(Minimum reading): Oxygen/Nitrogen: 0.0000001%(m/m)

Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.02%(m/m) or CV≦0.5% whichever is larger (Reference gas)
σn-1≦0.3%(m/m) or CV≦1.0%  whichever is larger (Standard sample)
** Based on the sample weight 1,0 ± 0,1g and with automation functions described below

Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD

Power of furnace
Iner gas fusion in impulse furnace
Enable to set the generating power 0 - 8,0kw

Sample introduction: Dual sample/flux introduction mechanism

Automation functions: Auto cleaner, Crucible loader

Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter

Sample ID: Enter up to 20 one-byte characters

Calibration
1) One point or multi point calibration (Reference gas or standard samples)
2) Calibration using previous analysis data
3) Calibration curve correction function

Functions
1) Display of realtime extraction curve
2) Analysis interruption
3) Self diagnosis and alarm display
4) Analysis of extraction curve (Superposition, difference etc analysis)
5) Out put (RC-232C or TCP/IP)

Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.

Weight: 230kg
On carrying, it is devided in two units (Under 140kg/unit)

Computer: PC model operating OS Windows® XP(SP2) and beyond

Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz

Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)

Ground resistance: Less than 10Ω

Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, Linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band

Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe(O.D.6mm) and suitable connecter fitting within 5m of unit

Cooling mechanism: Radiator unit (Separation type)

Electronic balance (Option)
Enable connection with electronic balance with 1-0,01mg minimum display

Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg

Schematics

EMGA-920 Dimensional Outline (mm)

Brochures

EMGA-920

Related Products

EMIA-8100 series

EMIA-8100 series equipped with electrical resistance furnace measures carbon and sulfur contained...

GD-Profiler

The GD-Profiler 2™ provides fast, simultaneous analysis of all elements of interest including the...

GD-Profiler HR

The HORIBA Scientific GD-Profiler HR™ gives the optimum in terms of resolution and number of...

Assistance tools with ICP-OES

ACTIVA-M, a bronze award winner for best new product at Pittcon 2007 as Master in multi-line...

SLFA-2100/2800

The SLFA-2100/2800 are designed specifically to meet the recent demanding needs of measuring the...

XGT-5200WR

High sensitivity XRF qualitative and quantitative analysis with an X-ray imaging...

XGT-7200

Single point analysis and automated hyperspectral imaging.

Dual vacuum modes.

Spot sizes from 1.2...

XGT-5200

Single point analysis and automated imaging.

Atmospheric pressure analysis.

Spot sizes from 1.2...

Other Available Models

EMGA-920AC: same performance as EMGA-920 but without crucible loader (equipped with automatic cleaner)
EMGA-920M: same performance as EMGA-920 but without crucible loader and without automatic cleaner (manual operation).