HO-300

Field-installation type ORP meter

HORIBA's long-selling water quality meter that excels in both indoor and outdoor applications.

Suitable for managing neutralization reaction processes such as aeration tanks and denitrification tanks. We offer a wide range of cleaning devices to accommodate a variety of samples.

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Comprehensive range of cleaning devices.

A diverse lineup of cleaning devices are offered.  Particularly, ultrasonic cleaners are advantageous as they do not require water or air, making them convenient for retrofitting existing HO-300 installations on site. The drop-in type of cleaning device, which fits into a holder, offers ease of maintenance.

Product nameORP meter
ModelHO-300
Combined sensorORP electrode
Measurable
range
ORP−2000 mV to 2000 mV (display range: −2200 mV to 2200 mV)
Temperature0℃ to 100℃
  • Display range :−10℃ to 110℃
    (the sensor types are detected automatically)
  • Display range :−20℃ to 130℃
    (the sensor types are input by manually)
Display
resolution
ORP1 mV
Temperature0.1℃
PerformanceORPRepeatabilityWithin ±5mV (response for equivalent input)
LinearityWithin ±5mV (response for equivalent input)
TemperatureRepeatability±0.3℃ (response for equivalent input)
Linearity±0.3℃ (response for equivalent input)
Transmission
output
Output type4 mA to 20 mA DC: input/output insulation type
Load resistanceMax.: 600Ω (case of 24 V DC power supply)*1
LinearityWithin ±0.08 mA (output only)
RepeatabilityWithin ±0.02 mA (output only)
Output rangeORP: Selection from preset ranges or free range input within measuring range.
Occasional out for errorHold or burnout to either 3.8 mA or 21 mA
Transmission holdIn the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact inputNumber of input points1
Contact typeNo-voltage a-contact
ConditionsON resistance: 40 Ω
Open voltage: 1.2 V
Short-circuit current: 21 mA DC max.
Contact functionExternal input for transmission holding.
Temperature
measurement
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature:
500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃)
Element selection methodAutomatic temperature sensor type detection or manual selection (no temperature element is selectable)
CalibrationORP adjust functionAdjust (offset adjustment): −200 mV to 200 mV
Span sensitivity correction: 0.500 to 1.500
Temperature calibration1 point calibration comparing reference thermometer
Self-checkCalibration errorTemperature calibration error
Out of temperature calibration range (for ORP sensor with RTD).
Electrode diagnostic errorReference electrode impedance error (only applicable for differential circuit mode with a liquid contact electrode)
Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range
Converter errorCPU, ADC, and memory errors
Operating temperature range−20℃ to 60℃ (without freeze)
Operating humidity rangeRelative humidity: 5% to 90% (without condensation)
Storage temperature−25℃ to 65℃
Power supplyRated voltage24 V DC (operating voltage range: 21 V to 32 V DC)*1
Power consumption0.6 W or less.
Compatible
standards
CE marking 
 Compatible standardsEMC directive: EN61326-1
RoHS Directive: EN50581
ImmunityIndustrial electromagnetic environment
Electostatic dischargeIEC61000-4-2
Electromagnetic field of radiated radio frequencyIEC61000-4-3*2
Electric fast transient/burstIEC61000-4-4
SurgeIEC61000-4-5*3
Conducted interference induced by radio frequencyIEC61000-4-6*2
EmissionClass A
Radiated disturbanceCICPR 11 CLASS A
FCC RulesPart15 CLASS A
StructureInstallationOutdoor installation type
Installation methodMounted on 50 A pole or wall
Protection codeIP65IEC60529, JIS C0920
Case materialAluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittingsSUS304
Material of hoodSUS304 (coated with epoxy modified melanine resin)
Material of windowPolycarbonate
Display elementReflective monochrome LCD
External dimensions180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets)
MassMain body: approx. 2.8 kg, Hood and mounting brackets: approx. 1 kg
  • *1:  The max. load resistance that can be connected is decided depending on the power-supply voltage.

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • *2:  The influence of the ORP reading by the electromagnetic field of radiated radio frequency or the conducted interference induced by radio frequency is within 6mV.
  • *3:  When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.

Note

Though an arrester (spark over voltage: 400 V) is implemented for transmission output, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.