Technical
| Weight | 122 lb, 55 kg. |
| Power supply | 100 V to 240 V (± 10%) - 50 Hz to 60 Hz |
| Dimensions | 22 in x 34.1 in x 23.3 in / 54 cm x 82 cm x 57 cm. |
| Printer | Laser |
| Throughout | 80 tests/hour. |
| Data management | Colour LCD touch screen: 12 in. |
| Memory | 10 000 results + graphics |
| Quality Control Management | 24 selectable QC files XB: 60 Operator selectable batches with statistics (20 samples per batch) With-in run Levey-Jennings Graphs |
| Logs | Reagents, Quality Control, Calibration, Maintenance, Blank Cycle, Patients Data, Settings, Information, Technology Communication, Errors, by date. |
| Patient reports Management | Delta check Anteriority (Matrix, curves, data) Manual input |
Samples
| 80 samples/hour (open or closed tube) | |
|---|---|
| Parameters (26) | WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*. |
| Graphs | High-definition LMNE matrix PLT, RBC, WBC/BAS curves. |
Blood Volume | CBC mode : 30 μL |
| Stability | 48 hours post-draw stability |
| Operating sample mode | Autoloader 10 racks (100 tubes) capacity |
| Identification | Alphanumeric. Bar-code reader (optional) |
* RUO parameters (Research Use Only)
Results
| Parameters | Presicion (cv %) | Linearity limits | Linearity Limits with CDR* |
|---|---|---|---|
| WBC | < 2.0 | 0-120 x 10³/µL | 120-360 x 10³/µL |
| RBC | < 2.0 | 0-8 x 106/µL | 0-8 X 106/µL |
| HGB | < 1.0 | 0-24 g/dl | 0-24 g/dL |
| HCT | < 2.0 | 0-67% | 0-67% |
| PLT (whole blood) | < 5.0 | 0-1900 x 10³/µL | 1900-3800 x 10³/µL |
| PLT (PLT concentrate) | < 5.0 | 0-2800 x 10³/µL | 2800-5600 x 10³/µL |














