Pentra ES 60

개요

  • PC를 통한 Data 관리
  • 전혈 micro-sampling (CBC : 30 µL - DIFF: 53 µL)
  • Cytochemistry, Impedance(실제 세포수를 측정) & Optical(광학도를 측정함에 따라 세포내부의 구조를 분석)-DHSS* Technology
  • 혈액 검체와 시약의 완벽한 혼합 - MDSS* Technology

기능

Pentra ES 60

  • 시간당 60검체 처리
  • 총 4개 시약 과 1개 희석제 장착
  • DHSS*기술로 5DIFF 를 완벽하게 분획
  • 5 DIFF Matrix와 함께 RBC, BAS/WBC, PLT 의 3 히스토그램을 보여줌
  • Basophil 별도 채널에서 측정
  • 비정형림프구(ALY***)와 미성숙세포(LIC***)는 높은 해상도의 Matrix에 분획으로 나타남(%, #)

* DHSS : Double Hydrodynamic Sequential System 
** MDSS : Multi-Distribution Sampling System
*** RUO parameters (Research Use Only)


제조원 HORIBA ABX SAS

사양

Technical

Weight

77 lbs / 35 kg

Power supply

100 to 240 V (± 10%) - 50 to 60 Hz.

Dimensions

20.3 x 17.5 x 19 in / 51,5 cm x 44,5 cm x 48 cm

Printer

Laser printer

Throughput

60 tests/hour.

Data management

Colour screen: 15 in. monitor (800 x 600px)
Windows XP Prof.TM 
User-defined flagging limits
Transmit patient & QC to LIS
Uni-directional or bi-directional connection
Communication Protocols: ABX / ASTM
CD-ROM drive / Floppy disk
2 serial ports minimum
USB Keyboard / USB Mouse

Memory

10,000 results + Graphics

Quality Control Management

12 selectable QC files
XB: 60 Operator selectable batches with statistics (20 samples per batch)
Within run
Levey-Jennings Graphs

Logs

Reagents, calibration, maintenance, errors, blank cycle

Samples

60 samples/hour (open tube)

Parameters (26)

WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*.

Graphs

High-definition LMNE matrix
PLT, RBC, WBC/BAS curves.

Blood Volume

CBC mode : 30 µL
CBC + DIFF mode : 53 µL

Stability

48 hours post-draw stability

Operating sample mode

Open tube

Identification

Alphanumeric. Bar-code reader (optional)

 * RUO parameters (Research Use Only)

Results

Parameters

Presicion (%CV)

Linearity

WBC

< 2.0

0-120 x 10³/µL

RBC

< 2.0

0-8 x 106/µL

HGB

< 1.0

0-24 g/dl

HCT

< 2.0

0-67%

PLT (whole blood)

< 5.0

0-1900 x 10³/µL

PLT (PLT concentrate)

< 5.0

0-2800 x 10³/µL

Related Documents