HF-960EM

線上感測器和自動量測切換濃度監測儀

符合半導體製程標準的氫氟酸濃度監測儀。 將感測器與完全 PFA(全氟烷氧基)接液零件結合在一起,尺寸優化至絕對極限。

它還可以處理高達 50% HF 濃度的寬範圍測量,滿足半導體製程的要求。

CONDChemical Concentration

DC24VTransmission OutputRS-485

事業部: Semiconductor
產品分類: Wet Process Control
製造商: HORIBA Advanced Techno Co., Ltd.

Wide-range Compatible

The auto-ranging feature, from low concentration to high concentration (up to 50% HF concentration), enables optimal measurement performance at each target concentration.

Contamination-free

Employing a sensor with a sanitary structure, using only PFA as the wetted material.  Being a chemically resistant sensor, it is capable of measuring various chemicals used in semiconductor processes.

High Stability/Reproducibility Accuracy

It achieves a reproducibility accuracy of ±0.5% FS.

Concentration Conversion Function

Concentration conversion is possible by inputting the relationship between chemical concentration and conductivity along with temperature characteristics. It is particularly suitable for the dilution management of low-concentration chemical solutions.

Specifications of HF-960EM
Model name HF-960EM
Measurement range HF: 0 - 50%, HCl: 0 - 10%, Conductivity: 0 - 2000 mS/cm
Repeatability +/- 0.5% F.S. (in equivalent input)
Linearity +/- 0.5% F.S. (in equivalent input)
Transmission output Number of output: 4
Contact output Outputs: 5 points
Contact input Inputs: 2 points
Communication output RS-485
Power supply DC 24 V +/- 10% 15W
Compatible sensor CM-520 Series
Wetted material of sensor PFA

Schematics

留言諮詢

如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。

* 這些欄位為必填項目。