Bench Meters


Touch panel realizes stress-free operation.
Intuitive, and very easy to use.
We’ve used our unrivalled experience together with detailed feedback from hundreds of customers to develop LAQUA, our most outstanding range of laboratory meters yet.
The HORIBA LAQUA series marries advanced technology with stylish design and an easy-to-use touch screen interface. The LAQUA series was developed as a durable meters offering long term reliability for laboratory professionals. Advanced functionalities and a wide array of electrodes provide high precision measurement with improved response time in a huge variety of applications.
LAQUA EC1100
Overview
Features
- Automatic and manual conductivity calibration up to 4 points
- Adjustable reference temperature and temperature coefficient
- Selectable cell constant and measurement unit
- Automatic / Manual temperature compensation and temperature calibration function
- Conductivity and total dissolved solids (TDS) conversion
- Four (4) TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
- Two (2) salinity calibration curves: NaCl, Seawater
- Built-in memory can store up to 500 data
- Auto-Hold / Auto-Stable measurement
- RS232 and USB outputs for connecting the meter to a printer or computer
- Security (password setting)
- 2 years meter warranty
Meter Kit Available
EC1100-S
The appearance of items in the picture are subject to change in each region.
Package Details:
- LAQUA EC1100 Conductivity/TDS/Res/Sal/Temp custom black and white static LCD meter
- Protection cover, integrated eletrode stand, power adaptor with 6 plugs (US, EU, UK, ANZ, China, Korea)
- 9382-10D plastic-body, Titanium/Platinum black conductivity probe (k=1.0) wth integrated temperature sensor
- 103-S: 84 uS/cm, 1413 uS/cm, 12.88 mS/cm, 111.8 mS/cm conductivity standard solutions
- Instruction manual
Manufactured by HORIBA Advanced Techno
Specifications
Model | LAQUA EC1100 Conductivity/TDS/Res/Sal/Temp |
---|---|
Condutivity Range | ...µS/cm to 2000 mS/cm |
Resolution | 0.05% F.S. |
Accuracy | ±0.6% F.S.; ±1.5% F.S. > 18.0 mS/cm |
Reference Temperature | 15 to 30 °C (adjustable) |
Temperature Coefficient | 0.00 to 10.00% (adjustable) |
Cell Constant | 0.1, 1.0, 10.0 (selectable) |
Measurement Unit | μS/cm, mS/cm, S/m (selectable / auto-ranging) |
Calibration Points | Up to 4 (Auto / Manual) |
TDS Range | 0.01 mg/L to 1000 g/L |
Resolution | 0.01 mg/L |
Accuracy | ±0.1% F.S. |
Calibration Curves | EN27888, 442, Linear (0.40 to 1.00), NaCl |
Resistivity Range | 0.00 Ω•cm to 200.0 MΩ•cm |
Resolution | 0.05% F.S. |
Accuracy | ±0.6% F.S.; ±1.5% F.S. > 1.80 MΩ•cm |
Salinity Range | 0.00 to 100.00 ppt / 0.000 to 10.000 % |
Resolution | 0.01 ppt / 0.001% |
Accuracy | ±0.2% F.S. |
Calibration Curves | NaCl / Seawater |
Temperature Range | -30.0 to 130.0°C |
Resolution | 0.1°C |
Accuracy | ± 0.4°C |
Calibration Option | Yes |
Memory Data | 500 |
Auto-Hold / Auto-Stable | Yes |
Cell Constant Display | Yes |
Electrode Status | On screen display |
Diagnostic Messages | Yes |
Display | Custom black and white static LCD |
Inputs | BNC, phono, DC sockets |
Output | RS232, USB |
Power Requirement | AC adaptor, 100-240V, 50/60Hz |
Weight | 500g |
Dimensions | 170(L) x 174(D) x 73(H) mm |
HORIBA Offices
Asia & Oceania
Singapore
HORIBA Instruments (Singapore) Pte Ltd.
work
83 Science Park Drive, #02-02A,
The Curie 118258 Singapore
work
+65 6 908 9660
work
fax
+65 6 745 8155
laqua@horiba.com
Europe, Middle East & Africa
Northampton
HORIBA UK Limited
Northampton Office
work
Kyoto Close,Summerhouse Road,
Moulton Park
Northampton NN3 6FL UK
Moulton Park
work
+(44) 1604 542 600
work
fax
+(44) 1604 542 699
waterquality@horiba.com