Combustion gas appliances evaluation and inspection
Gas-burning apparatuses such as gas water heaters and gas stoves have safety standards, and in order to make them pass tests of conformity to such standards, it is important to confirm their safety performance in the research and development stage. The PG-300 Series portable gas analyzers offer a full array of options that meet different types of sample gas, enabling more precise analysis of gas.
Sample option combination for combustion gas appliances
Drain separator unit
Drain pot unit
When the gas sample includes moisture ranging from ambient temperature saturation to 40 °C saturation, a Drain Separator and Drain Pot are installed at the stage before the analyzer unit.
- Drain separator
- Drain pot
■Drain separator unit / Drain pot unit specifications
Model | DS-300 | DP-300 | |
---|---|---|---|
Sample conditions | Temperature | 0°C to 40°C / 32°F to 104°F | |
Moisture | Ambient temperature saturation ~ 40°C / 104°F saturation | ||
Dust | 0.1 g/m3 or less | ||
Pressure | ±0.98 kPa | ±4.9 kPa |
Electronic cooler unit
When the gas sample includes moisture exceeding 40 °C saturation, or when conducting continuous measurement (for five days or less), a thermoelectric cooler is installed at the stage before the analyzer unit.
The electronic cooler unit can also accommodate low-temperature SO2 measurements.

■Electronic Cooler unit specifications
Model | PS-300 |
---|---|
Material in contact with gas | Ti, SUS, PVC, PTFE, FKM, PVDF, PP, Glass |
Inlet sample | Approx. 2 L/min. |
Dehumidify | 15°C / 59°F saturated |
Usable temperature | 0°C to 40°C / 32°F to 104°F |
Usable humidity | 85% or less |
Power | 100/110/115/120/220/230/240 V AC, 50 Hz/60 Hz (depend on specifications) |
Outline | 260(W) x 375(D) x 235(H) mm / 10.2"(W) x 14.7"(D) x 9.2"(H) (except for protrusion) |
Mass | Approx. 12kg / 26.46 lb |
Sample gas condition | Temperature: Ambient temperature, Dust: 0.1g/m3 or less, |
[Halogen scrubber] (optional)
The Halogen scrubber can be built into the electronic cooling unit as an option. It is used to prevent corrosion of the cells, tubes and other internal components when the gas analyzer is operated at waste incineration facilities or in other situations where the gas sample includes Cl2.
Primary side filter probe
Either of two types may be selected depending on use.
- Flue probe
- Simple probe
■Primary side filter probe specifications
Model | Simple probe | SE3 (flue probe) | |
---|---|---|---|
Probe length (standard) | 10 cm / 3.937" | 1 m / 3'28" | |
Sample conditions | Temperature | 0°C to 50°C / 32°F to 122°F* | 0°C to 120°C / 32°F to 248°F |
Moisture | 40 vol% or less | ||
Dust | 0.1 g/m3 or less | ||
Pressure | ±2.94 kPa |
*At flange inlet
Note:
- Please contact HORIBA if the analyzer will be used in environments in which the temperature exceeds 120 °C.
- Please contact HORIBA in case of use under the environmental that the pressure condition is other than ±2.94 kPa.
Product information
The PG-300 is a compact and lightweight stack gas analyzer that can simultaneously measure up to five separate gas components...