Industry
ABX Pentra DF 120
Overview
- Autoloader with continuous loading capability
- Double DIFF Matrix
- Automatic reflex testing
- Validation station with cytology atlas
- DHSS technology
Features
ABX Pentra DF 120
- Methods of measurement: Cytochemistry, impedance, absorbance and flow cytometry
- Balance Concept: Automatic control of the leukocyte count based on 3 independent channels
- DHSS: Focused flow cytometry and sequential measurement (impedance and absorbance)
- Automatic re-run of samples: Selective and programmable (hematology parameters alarms and flags)
- Rotative mixing of samples: Smooth and efficient
- Liquid valves ABX (no pinch valves): Reliability and volume accuracy
Manufactured by HORIBA ABX SAS
Specifications
Technical
Weight | 170 kg |
Power supply | 100/240 V - 50/60 Hz - 700 W |
Dimensions | 75 x 167 x 55 (H x W x D) cm |
Printer | Laser printer |
Throughout | 120 tests/hour |
Data processing | OS9 |
Compliance | ASTM protocol |
Memory capacity | 90 000 results + graphics |
Samples
120 tests/hour | |
---|---|
Parameters (32) | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV,PCT*, PDW*, NEU, LYM, MON, EOS, BAS, ALY, LIC, IMM, IMG, IML: (% and #) |
Graphs | High-definition DIFF matrix, PLT, RBC, WBC/BAS curves |
Volume | Manual mode : 130 µL |
Stability | Up to 48 hours for CBC/DIFF |
Operating sample mode | Autoloader 15 racks (150 tubes) capacity |
Identification | Alphanumeric, Bar-code reader (internal and external) |
Results
Parameters | Presicion (cv %) | Linearity |
---|---|---|
WBC | < 2% | 0-150 10³ / mm³ |
RBC | < 2% | 0-8 106 / mm³ |
HGB | < 1% | 0-24 g/dl |
HCT | < 2% | 0-67% |
PLT (whole blood) | < 5% | 0-1900 10³ / mm³ |
PLT (PLT concentrate) | < 5% | 0-2800 10³ / mm³ |
LYM | < 5% | -- |
MON | < 15% | -- |
NEU | < 3% | -- |
EOS | < 25% | -- |
BAS | < 40% | -- |