Turbidity Probe

Our Turbidity probes are based on Nephelometric Method (ISO 7027 – EN 27027). A light beam is sent to the sample through an optical lens. The 90 degree scattered light by suspended particle is collected by the probe through a second lens and it is converted in an electric signal proportional to the turbidity of the sample.
The probe uses an infrared light and the measuring is not affected by the color of the sample. They are suitable for measurement of turbidity in water quality monitoring, municipal and industrial water treatment and Aquaculture.

Application

  • Industrial waste water treatment plant
  • Drinking water treatment plant
  • Other process water application
  • Municipal waste water treatment plant
  • Surface water application
Dimension DrawingMeasurement Range0 – 4.000 NTU, 0 – 40.00 NTU, 0 – 400.0 NTU
Measurement TechniqueNephelometric
Detector90° scattered light detector
Light SourceInfrared
Ambient Temperature Range-5 to 50 °C
Working Pressure6 bar maximum at 25 °C
Body materialPVC
Integral Sensor Cable Length10 meter
MountingIn-line or In-flow
Probe warranty1 Year

 

Reference part

EXT-TURBNEPH-HNephelometric turbidity probes high range
Range: 0 – 400 NTU
10 meters cable
EXT-TURBNEPH-MNephelometric turbidity probes medium range
Range: 0 – 40 NTU
10 meters cable
EXT-TURBNEPH-LNephelometric turbidity probes low range
Range: 0 – 4 NTU
10 meters cable

Informationsanfrage

Sie haben Fragen oder Wünsche? Nutzen Sie dieses Formular, um mit unseren Spezialisten in Kontakt zu treten.

* Diese Felder sind Pflichtfelder.