XS Analyzer - Multi-Gas CEMS
Overview
The HORIBA XS analyzer utilizes chemiluminescence and magnetopneumatic as well as cross-flow modulated no-dispersive infrared (NDIR) technologies, providing highly accurate measurement for up to five gasses plus system control, yet requiring no routine maintenance nor optical alignment.
The XS analyzer features specialized designs such as the optimized path length and high sensitivity detector that give precision measurements with a sample flow rate of less than one liter per minute.
The XS analyzer automatically initiates system calibration at intervals predefined by the user. The easy-to-read display can show concentrations for any of up to five gasses, or display averages or diluent-corrected values.
A micro-processor provides true computerized control plus self-diagnostics that allow the user to easily determine the source of an alarm or caution signal displayed on the front panel.
Features
- Measurement of up to five gases.
- Highly accurate measurements.
- No optical alignment required.
- Field-proven technologies.
- No routine maintenance required.
Manufactured by HORIBA
Specifications
Specification | Specification value |
|---|---|
Measured Components | One to five of the following gases: NOx, SO2, CO, CO2 and O2 |
Ranges available NOX CO SO2 CO2 O2 |
0-100/5000 ppm 0-50/5000 ppm standard or 0-10/1000 ppm optional 0-50/5000 ppm standard 0-5/50 % 0-10/25 % |
Repeatability Ranges>/= 200 ppm Ranges</= 200 ppm |
± 0.5% of full scale ± 1.0% of full scale |
Response time NO, CO, and CO2 SO2 and O2 |
15 seconds for 90% response 30 seconds for 90% response |
Drift Zero drift Ranges> 200 ppm Ranges< 200 ppm Span drift | For ambient temperature within ±10°F (5°C)
±1.0% of full scale / 7 days ±2.0% of full scale / 7 days ±2.0% of full scale / 7 days |
Linearity | ±1.0% of full scale |
Interference | <±2.0% of full scale for standard sample gas composition |
Power requirements | 115 v ac, 50 / 60 Hz |
Power consumption | 300 VA |
Sample gas flow rate | 0.5 lpm |
Sample gas pressure | 4.9 kPa (500 mm Aq) ±0.5% |
Outputs | 0-16 mA or 4-20 mA and 0-1 V DCisolated output for up to 10 channels and 16 discrete outputs available rated 1A, 250V AC or 125 V DC resistive load. |
Materials exposed to sample gas | 304 stainless steel TeflonTM, polypropylene, fluororubber, PVC, and glass |
Dimensions: Height: Width: Length: |
5 ¼ in (134 mm) 19 in (483 mm) 20 in (508 mm) |
Weight | Approx. 34 lb (15.4 kg) |