Technical Specifications
STAT sample priority
Throughput: Up to 200 tests/hour, up to 400 tests/hour with ISE
Reagent/Sample Handling
Reagent/Sample tray: 40 positions for reagents and 40 positions, Internal bar code reader
Reagent/Sample probe: Liquid level detection, vertical collision protection and inventory checking, reagent pre-warming
Reaction System
Cuvette: 40 disposable cuvettes
Reaction volume: 100µ1 ~360µ1
ISE Module (optional)
Measuring K+, Na+, Cl-
Optical System
Light Source: Halogen-tungsten lamp
Wavelength: 8 wavelengths, 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm
Control and Calibration
Calibration modes: Linear (one point, two points and multi-points), Logit-Log 4P, Logit-Log SP, spline, exponential, polynomial, parabola
Control Rules: X-R, L-J, Westgard multi-rule, Cumulative sum check, twin plot