HO-300

Field-installation type ORP meter

HORIBA's long-selling water quality meter that excels in both indoor and outdoor applications.

Suitable for managing neutralization reaction processes such as aeration tanks and denitrification tanks. We offer a wide range of cleaning devices to accommodate a variety of samples.

Abteilung: Process Water
Produktionsfirma: HORIBA Advanced Techno Co., Ltd.

Comprehensive range of cleaning devices.

A diverse lineup of cleaning devices are offered.  Particularly, ultrasonic cleaners are advantageous as they do not require water or air, making them convenient for retrofitting existing HO-300 installations on site. The drop-in type of cleaning device, which fits into a holder, offers ease of maintenance.

Product name ORP meter
Model HO-300
Combined sensor ORP electrode
Measurable
range
ORP −2000 mV to 2000 mV (display range: −2200 mV to 2200 mV)
Temperature 0℃ to 100℃
  • Display range :−10℃ to 110℃
    (the sensor types are detected automatically)
  • Display range :−20℃ to 130℃
    (the sensor types are input by manually)
Display
resolution
ORP 1 mV
Temperature 0.1℃
Performance ORP Repeatability Within ±5mV (response for equivalent input)
Linearity Within ±5mV (response for equivalent input)
Temperature Repeatability ±0.3℃ (response for equivalent input)
Linearity ±0.3℃ (response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC: input/output insulation type
Load resistance Max.: 600Ω (case of 24 V DC power supply)*1
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range ORP: Selection from preset ranges or free range input within measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact input Number of input points 1
Contact type No-voltage a-contact
Conditions ON resistance: 40 Ω
Open voltage: 1.2 V
Short-circuit current: 21 mA DC max.
Contact function External input for transmission holding.
Temperature
measurement
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature:
500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃)
Element selection method Automatic temperature sensor type detection or manual selection (no temperature element is selectable)
Calibration ORP adjust function Adjust (offset adjustment): −200 mV to 200 mV
Span sensitivity correction: 0.500 to 1.500
Temperature calibration 1 point calibration comparing reference thermometer
Self-check Calibration error Temperature calibration error
Out of temperature calibration range (for ORP sensor with RTD).
Electrode diagnostic error Reference electrode impedance error (only applicable for differential circuit mode with a liquid contact electrode)
Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range
Converter error CPU, ADC, and memory errors
Operating temperature range −20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature −25℃ to 65℃
Power supply Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC)*1
Power consumption 0.6 W or less.
Compatible
standards
CE marking  
  Compatible standards EMC directive: EN61326-1
RoHS Directive: EN50581
Immunity Industrial electromagnetic environment
Electostatic discharge IEC61000-4-2
Electromagnetic field of radiated radio frequency IEC61000-4-3*2
Electric fast transient/burst IEC61000-4-4
Surge IEC61000-4-5*3
Conducted interference induced by radio frequency IEC61000-4-6*2
Emission Class A
Radiated disturbance CICPR 11 CLASS A
FCC Rules Part15 CLASS A
Structure Installation Outdoor installation type
Installation method Mounted on 50 A pole or wall
Protection code IP65 IEC60529, JIS C0920
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 (coated with epoxy modified melanine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets)
Mass Main body: approx. 2.8 kg, Hood and mounting brackets: approx. 1 kg
  • *1:  The max. load resistance that can be connected is decided depending on the power-supply voltage.

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • *2:  The influence of the ORP reading by the electromagnetic field of radiated radio frequency or the conducted interference induced by radio frequency is within 6mV.
  • *3:  When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.

Note

Though an arrester (spark over voltage: 400 V) is implemented for transmission output, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.