HE-200C

Field-installation type electric conductivity meter (conductivity meter)

For measurements in applications such as pure water, boiler water, and pharmaceutical manufacturing processes.

Suitable for measurements in applications such as pure water, boiler water, and pharmaceutical manufacturing processes. Our strength lies in our sensor lineup, suitable for a variety of samples, including sensors for pharmaceutical manufacturing processes that comply with the pharmacopoeias (USP, EP, JP) of various countries, and in our robustness to withstand harsh environmental conditions.

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

A lineup of sensors to match sample properties and applications

We offer general-purpose sensors suitable for pure water and boiler water control and wastewater treatment monitoring, as well as sensors for pharmaceutical manufacturing processes that comply with the pharmacopoeias of various countries (USP, EP, JP).

Highly robust

The accessories and wiring required for installation have the robustness typical of industrial applications, allowing them to withstand harsh environmental conditions.

Product name Conductivity meter
Model HE-200C
Sensors in combination 2-electrode method conductivity sensor
ESH, ESD, ESL, and FS Series sensor of cell constant 0.01/cm,0.1/cm, or 1.0/cm
Measurement
range
Cell constant (/cm) 0.01
(ESH/ESD/ESL-001)
0.1
(ESH/ESD/ESL/FS-01)

1.0
(ESH/ESD/ESL-1)

  Unit
Conductivity*1 μS/cm 0.000 to 2.000
0.00 to 20.00
0.000 to 2.000
0.00 to 20.00
0.0 to 200.0
0 to 2000*2
0.0 to 200.0
0 to 2000
mS/m 0.0000 to 0.2000
0.000 to 2.000

0.0000 to 0.2000
0.000 to 2.000
0.00 to 20.00
0.0 to 200.0*2

0.00 to 20.00
0.0 to 200.0
TDS
conversion*1
mg/L 0.00 to 2.00
0.0 to 20.0
0.00 to 2.00
0.0 to 20.0
0 to 200
0 to 200
0 to 2000
Temperature 0 to 100 (display range: −10℃ to 160℃)
Display
resolution
Conductivity/TDS conversion As shown in the measuring table.
Temperature 0.01℃
Performance Conductivity Repeatability Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/
m for FS-01 sensor
Linearity Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor
TDS
conversion
Repeatability Within ±1.5% of full scale value (response for equivalent input)
Linearity Within ±1.5% of full scale value (response for equivalent input)
Temperature Repeatability ±0.1℃ (response for equivalent input)
Linearity 0.5℃ (response for equivalent input)
Transmission
output
Number of output 2 (The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Max.: 900Ω
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Output 1 Conductivity: Free setting within a measuring range.
Output 2 Temperature: Free setting within a range between −10℃ and 160℃
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact output Number of output 3
Output type No-voltage contact output
Contact type Relay contact; SPDT (1c)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact
function
R1 and R2 Selectable from upper limit alarm, lower limit alarm, USP conductivity requirement, and currently holding of transmission output.
(The contact is closed during alarm operation, opened normally and while the power is down.)
FAIL Error alarm
(Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting range Setting range: within a measuring range
Delay time: 0 seconds to 600 seconds
Contact input Number of input points 1
Contact type Open collector non-voltage a contact
Conditions ON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12mA DC max.
Contact function External input for transmission holding.
Communication
function
Communication type RS-485
Signal type 2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Temperature element Platinum resistor: 1 kΩ (0℃)
Temperature compensation
  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95 ℃)
  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95 ℃; temperature coefficient: ±3%/℃)
  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)
Temperature compensation
range
0 ℃ to 100 ℃(However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Conductivity Based on parameter input of coefficient for the sensor cell constant.
TDS conversion Conversion by user-defined coefficient value (TDS calculation is 0.30 to 1.00 times of sample conductivity by μS/cm.)
Self-check Sensor diagnosis error Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range
Meter error CPU error, ADC error, and memory error
Operating temperature range −20℃ to 55℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature −25℃ to 65℃
Power supply Rated power supply voltage 100 V to 240 V AC ±10% 50/60 Hz
Power consumption 15 VA (max.)
Others With power switch for maintenance use
Compatible
standards
CE marking EMC   :EN61326-1
      Class A, Industrial electromagnetic environment
Safety :EN61010-1
RoHS  :EN50581
    9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Mount on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 3.5 kg
Hood and mounting brackets: Approx. 1 kg

*1:  Only the sanitary sensor (FS-01 series) is applied.
*2:  Conductivity measurement and TDS conversion measurement cannot be selected at the same time.
*3:  When the sensor cable, the transmission cable, or the contact input cable is extended to 30 m or longer, the surge test specified in the EMC directive for CE marking is not applied.

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.