HE-200H

Field-installation type electric conductivity meter

For measurement of electrical conductivity in aqueous solution, concentration adjustment and rinse monitoring for CIP, and process concentration gauge. densitometry.

Appropriate for concentration adjustment and rinse monitoring for CIP, wastewater treatment, seawater and hydroponic solution monitoring, as well as a process concentration gauge. Our sensor lineup offers options suitable for various sample types, and their strength lie in their robustness to withstand harsh environmental conditions.

COND

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

A lineup of sensors to match sample properties and applications

Two types of sensors are available: a lead type which can be directly thrown into the tank or placed in the tank with an immersion holder, and a connector type for direct insertion into piping or with a Flow-through holder.

Highly robust

Accessories required for installation and wiring possess the robustness appropriate for industrial applications, allowing them to withstand harsh environmental conditions.

Salinity concentration conversion function

The transmitter is equipped with a salinity concentration conversion function that is suitable for monitoring the salinity of seawater.

Product name Conductivity meter
Model HE-200H
Sensors in combination 4-electrode method conductivity sensor (FES Series) of cell constant 0.1/cm, or 1.0/cm
Measuring
range
Measurement
type
Unit Cell constant
(/cm)
0.1
(FES-2XX, 310 series)
1.0
(FES-125, 126)
Conductivity mS/cm 0.000 to 2.000
0.00 to 20.00
0.0 to 200.0
0 to 2000 △(*Note2) △(*Note2)
Auto (*Note1)
Practical region
(*Note2)
0 to 500 0 to 1000
Display range 0 to 2200
S/m 0.000 to 0.2000
0.00 to 2.000
0.00 to 20.00
0.0 to 200.0 △(*Note2) △(*Note2)
Auto (*Note1)
Practical region
(*Note2)
0.0 to 50.0 0.0 to 100.0
Display range 0.0 to 220.0
Temperature 0℃ to 100℃ (display range: −30℃ to 160℃)
Salinity
conversion
capability
Seawater 0.00% to 4.00%
NaCl 0.0% to 20.0%
Concentration
conversion
capability
NaOH 0.00% to 5.00%
HNO3 0.00% to 5.00%
H3PO4 0.00% to 5.00%
Option 1 to 4 0.00% to 100.00%
Display
resolution
Conductivity
Conversion density
As shown in the measuring and conversion table.
Temperature 0.01℃
Performance Conductivity Repeatability
(mS/cm)
Cell constant 0.1 (/cm) 1.0 (/cm)
0.00 to 20.00 Within ±0.5% of the full scale Within ±0.5% of the full scale
20.0 to 200.0 Within ±1.0% of the full scale Within ±0.5% of the full scale
200 to 1000 Within ±1.0% of the full scale
Condition response for equivalent input
Linearity
(mS/cm)
Cell constant 0.1 (/cm) 1.0 (/cm)
0.00 to 20.00 Within ±0.5% of the full scale Within ±0.5% of the full scale
20.0 to 200.0 Within ±1.0% of the full scale Within ±0.5% of the full scale
200 to 1000 Within ±1.0% of the full scale
Condition response for equivalent input
Performance Temperature Repeatability ±0.1℃ (response for equivalent input)
Linearity ±0.5℃ (response for equivalent input)
Transmission
output
Number of output 2 (The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 900Ω
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Output 1 Conductivity or density converted: Free setting within a measuring range.
Output 2 Temperature: Free setting within a range between −30℃ and 160℃
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact
output
Number of output 3
Output type No-voltage contact output
Contact type Relay contact; SPDT (1c)
Contact capacity AC 250 V 3 A, DC 30 V 3 A
Contact
function
R1 and R2 Selectable from upper limit alarm, lower limit alarm, and currently holding of transmission output.
(The contact is closed during alarm operation, opened normally and while the power is down.)
FAIL Error alarm
(Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting range Selectable from conductivity and density
  • Setting range: Within a measuring range
  • Delay time: 0 to 600 seconds
Contact input Number of input 2
Contact type No-voltage "a" contact for open collector
Conditions ON resistance: 100Ω max.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Function of input The input signal use can be changed.
Two inputs for four output current range selection, or, one input for two current range selection and one input for holding.
The display also responds for the range selection by input.
Communication function Communication type RS-485
Signal type 2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Temperature element Platinum resistor: 1 kΩ (0℃)
Compensation method
  • Temperature characteristics of NaCl (reference temperature: 5℃ to 95℃)
  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: ±3%/℃)
  • No temperature compensation
    One of the three ways of compensation is selectable.
Temperature compensation
range
0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)
Calibration Conductivity Based on parameter input of coefficient for the sensor cell constant.
Self-check Sensor diagnosis error Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range
Converter CPU error, ADC error, and memory error
Operating temperature range −20℃ to 55℃ (without freeze)
Operating humidity range Relative humidity: 5℃ to 90℃ (without condensation)
Storage temperature −25℃ to 65℃
Power supply Rated power supply voltage 100 V to 240 V AC ±10% 50/60 Hz
Power consumption 15 VA (max)
Others With power switch for maintenance use
Compatible
standards
CE marking EMC    :EN61326-1
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS   :EN50581
     9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Mount on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of food SUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) mm x 155 (H) mm x 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg

Note

  • Note 1  : The decimal point position changes automatically
  • Note 2  : The actual limit of measuring range of raw conductivity is 500 mS/cm for 0.1/cm and 2000 mS/cm for 1/cm sensor. (The sensor durability for chemicals is not assumed.)
  • Note 3  : When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 4  : An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.