Yumizen C600

Automated Clinical Chemistry Analyzer

This solution is specifically designed for mid to high- end laboratories seeking superior performance combined with user-friendly operations.

It offers an impressive throughput of 400 photometric tests per hour, which increases to 600 tests with ISE integration, ensuring prompt and precise diagnostic results.

Its high-speed processing capabilities maintain an uninterrupted workflow, even during peak operational hours, effectively minimizing manual intervention and delays.

The system supports automated sample loading, reagent dispensing, and result analysis, delivering a seamless and efficient laboratory experience.

Segment: Healthcare
Division: Clinical Chemistry
Manufacturing Company: HORIBA India Private Limited
  • 400 photometric tests/hour & 600 with ISE
  • Continuous Sample Loading with barcode scanning
  • Maintenance-Free Ceramic pistons ensuring consistent and accurate pipetting
  • MALS (Multi Array LED System) technology with 8 LED wavelength power
  • Separate heavy duty dispensing probes with liquid level & Clot detector
  • On board laundry system designed with hot water for efficient cleaning

Technical Specifications

Throughput 400 tests/hr (Photometric)320 tests/hr (ISE)600 tests/hr (Photometric + ISE)
Principles of analysis Colorimetry, Turbidimetry, Potentiometry
ISE Module (optional)  
Sample type Serum, Plasma or Urine
Electrode type Na⁺, K⁺, Cl⁻, Li⁺ (optional)
Sample volume Serum: 100 µL, Urine: 200 µL
Sample Handling  
Capacity of sample tray 126 (up to 9 segments – 14 positions/segment)
Sample Barcode Reader Yes
Positions with barcode 81
Primary Tube Size Diameter from 12 mm to 16 mm (height up to 100 mm)
Sample containing type Vacutainer, 2ml cup, 500µL Pediatric Cup, Test tubes, etc.
Type of syringe Ceramic piston pump with low maintenance
Sample volume From 2 µL to 40 µL
Predilution ratio From 1:2 to 1:200
Level detector Yes
Clot detector Yes
Vertical collision detector Yes
Optical System  
Light Source LEDs
Photometric range -0.2 A to 3.5 A
Wavelengths 340, 405, 505, 535, 560, 600, 635, 670 nm
Filters bandwidth ± 2 nm
Internal resolution 0.0001
Photo Detector Main Photodiode + Reference Photodiode
Measurement precision CV <1% to 0.1 A CV <0.1% to 2 A
Environmental Requirements  
Ambient temperature From 10 °C to 35 °C From 10 °C to 30 °C (with ISE module)
Relative humidity <85% without condensation
Altitude <2500 m
Physical Information  
Dimensions 1200 mm(W) x 720 mm(D) x 1258 mm(H)
Weight: 210 Kg
Electrical Requirements  
Mains voltage 115 V to 230 V
Mains frequency 50 Hz or 60 Hz
Electric power 500 VA
Fluidic Requirements  
Water inlet External tank or Direct DI water supply
Water Type Purified Type II
Water consumption <14 liters/hr
Container for biohazard waste 5L
Container for washing solution 5L
Minimum Computer Requirements  
Operating system Windows 10 or above, 64-bit (x64)
CPU Equivalent to Intel Core i3
RAM 8 GB or higher
Hard Disk 500 GB or higher
Monitor minimum resolution Minimum resolution 1024 x 768
Connector of serial channel USB