HC-200F

Field-installation Type Fluoride Ion Concentration Meter

For the management of wastewater from semiconductor/FPD plants and glass factories.

A fluoride ion concentration meter that covers a wide range and offers a diverse lineup of cleaning devices. Combined with the fluoride ion electrode 1009 to detect free fluoride ions in the sample. Ideal for managing wastewater from semiconductor/FPD factories and glass factories.

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Supporting a wide range

With a single unit, a wide range from low to high concentrations can be covered.

Comprehensive range of cleaning devices.

A diverse lineup of cleaning devices are offered.

Product name Fluoride ion meter
Model HC-200F
Electrode used with product Fluoride ion electrode
Measurement
range
Concentration 0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L)
Available range options:
10000, 5000, 2000, 1000, 500, 200, 100, 50, 20.0, 10.0 mg/L
Temperature 0℃ to 100℃ (Display range: -20℃ to 130℃)
Display
resolution
Concentration 0.1 mg/L: 0.0 mg/L to 20.0 mg/L
1 mg/L: 0 mg/L to 200 mg/L
10 mg/L: 0 mg/L to 2000 mg/L
100 mg/L: 0 mg/L to 10000 mg/L
Temperature 0.1℃
Performance Concentration Repeatability Within ±7% of the full scale (response for equivalent input)
Linearity Within ±10% full scale (response for equivalent input)
Temperature Repeatability ±0.3℃ (response for equivalent input)
Linearity ±0.3℃ (response for equivalent input)
Transmission
output
Number of output points 2 (the negative terminals for transmission outputs are internally connected and at the same electric potential)
Output type 4 mA to 20 mA DC, input/output isolated type
Load resistance Max.: 900Ω
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Output 1 Concentration: Selection from preset ranges or free range input within measuring range.
Output 2 Temperature: Free setting within a range between -20℃ and 130℃
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact output Number of output points 3 points
Output type No-voltage contact output
Contact type Relay contact, SPDT (1c)
Output capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact
function
R1, R2 Selectable from upper limit alarm, lower limit alarm, ON/OFF control,currently holding of transmission output, and cleaning output (The contact is closed during alarm operation, opened normally and while the power is down).
FAIL Error alarm (Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting range
  • Setting range: within a measuring range
  • Delay time: 0 seconds to 600 seconds
Control setting range
  • Setting range: within a measuring range
  • Control width: 2 mg/L to 40% of the setting range
Cleaning output Number of output points 1
Output type Contact output with voltage (output of connected power supply voltage)
Contact type Relay contact; SPST (1a)
Contact capacity 250 V AC 0.5 A
Contact function Actuation of solenoid valve for cleaning
Settings Cleaning period 0.1 hours to 168.0 hours
Cleaning time 2 seconds to 600 seconds
Hold time 2 seconds to 600 seconds
Timer accuracy Within 2 minutes per month
Description of cleaning
operation
  • Operation of internal timer
  • Operation of internal timer and external input
  • Enable the internal timer only when external input is active.
  • Choose one of the cleaning start signals (when this signal is ON for 2 seconds minimum, the internal cleaning sequence starts).
Contact input Number of input points 1
Contact type No-voltage “a” contact of open collector
Conditions ON resistance: 100Ωmax.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact function External input for cleaning or transmission holding if cleaner is not attached.
Transmission
capability
Communication type RS-485
Signal type 2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Applicable temperature element Platinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature: 10 kΩ (25℃)
measurement range of temperature 0℃ to 100℃
Temperature calibration 1 point calibration comparing reference thermometer
Calibration Number of calibration points Selectable from 1, and 2 points
Kinds of standard solutions The first point: Calibrated by the standard solution with the density from 50% to 100% of the measuring range.
The second point: Calibrated by the standard solution with the density from 1% to 20% of the measuring range.
Additional capabilities Automatic detection of calibration failure (asymmetry potential,sensitivity, or response time)
Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration)
Self-check Calibration error Asymmetry potential error, and sensitivity error
Temperature calibration error
Electrode diagnostic error Temperature sensor short-circuit, temperature sensor disconnection,and out of the temperature measurement range
Converter error CPU error, ADC error, and memory errors
Operating temperature range -20℃ to 55℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature -25℃ to 65℃
Power supply Rated power supply voltage 100 V to 240 V AC ±10% 50/60 Hz
Power consumption 15 VA (max)
Others With power switch for maintenance use
Compatible
standards
CE marking EMC  :EN61326-1
    Class A, Indrial electromagnetic environmentust
Safety :EN61010-1
RoHS  :EN50581
    9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Mounted on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS 304 stainless steel
(coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 3.5 kg
Cover and mounting brackets: Approx. 1 kg

*:  When the sensor cable, the transmission cable, or the contact input cable is extended to 30 m or longer, the surge test specified in the EMC directive for CE marking is not applied.

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.