ENDA-C2000 Stack Gas Analyzer System
概要
The ENDA-C2000 Series is a lineup of units for the continuous monitoring of residual NH3 contained in emissions from ammonia dry method denitrification equipment. Measurement is highly accurate thanks to the use of a new drift-free chemiluminescent analysis (CLA) device. Used in semiconductor fabrication plants, the units are designed to provide a long life. The units are simple to use thanks to such features as an integrated analysis and operating instruments, front-panel operation, and self-diagnosis. The units are well suited for denitration catalyst inspection, controlling NH3 injection volumes, the prevention of ammonium sulfate crystal formation, facilities inspection, and process control.
特征
The units are equipped with a new drift-free chemiluminescent analysis (CLA) device.
- The units are capable of a minimum range of 0ppm-10ppm for NH3 and NOx.
- The units are equipped with a revolutionary magnetopneumatic method oxygen analysis device that works using oxygen's ordinary magnetism.
- Front-panel operation offers enhanced ease of use.
- Equipped with numerous functions, such as computer monitoring of analysis conditions, control, data processing, and memory, the units are simple to use and require minimal maintenance.
- The units are fully equipped with self-diagnostic functions.
- The units can be configured to meet the needs of a wide range of installation sites.
制造: HORIBA
規格
MEASURING COMPONENT & RANGE
NH3 |
|
|---|---|
NOX |
|
O2 |
|
BASIC PERFORMANCE (-5°C to 40°C)
Ambient Temperature | -5°C to 40°C | |
|---|---|---|
Repeatability | ±0.5 FS% (±1.0 FS% when optinal range is included) | |
Drift(NOx analyzer) | Zero drift: ±1.0 FS%/week Span drift: ±2.0 FS%/week Provided the ambient temperature change is ±5°C | |
Total interference | ±2.0 FS% | |
BASIC PERFORMANCE (40°C to 50°C)
Ambient Temperature | 40°C to 50°C | |
|---|---|---|
Repeatability | ±1.0 FS% (±2.0 FS% when optional range is included) | |
Drift(NOx analyzer) | Zero drift: ±2.0 FS%/week Span drift: ±2.0 FS%/week Provided the ambient temperature change is ±5 °C | |
Total interference | ±3.0 FS% | |
SYSTEM SPECIFICATION
Response speed | Td + T90 = 90 scc. Max. from analyzer inlet Td + T90 = 70 scc. Max. from calibration gas inlet | ||
|---|---|---|---|
Supply voltage | AC 100 V ± 15% (85 V–115 V) | ||
Power ferquency | 47.5 Hz–63.0 Hz (common frequency) | ||
Sampling method | Dehumidified sampling at 2.5 °C (2 staged electronic clling unit, depressurized sampling) | ||
Sample gas flow | 1.5 L/min for each line | ||
Power consumption | 1.9 kVA, Separate 600 VA for cold district specification | ||
Analog output | 1–3 systems selected in combination of any one of DC4–20 mA/0–16 mA/1–5V and DC 0–1 V. lsolation type | ||
Stadus signals | Analyzer error, calibraion error, power failure (or NFB tripped), range indication, calibraiing, under maintenance, upper/lower limit alam (optional), purging(optional), contact capacity DC 125 V, 1A, AC 250 V, 1A resistance load | ||
OUtside demensions | (W) 1000 mm × (H) 1800 mm × (D) 1050 mm | ||
Mass | Approx. 500 kg (gas cylinder excluded) | ||