HE-300C

현장 설치형 전기전도도계 (2-Wire Transmission, 저농도용)

HE-300C는 전도도 센서(ESH/FS Series)를 연결하여 시료수의 전도도 및 온도를 측정합니다.
품질 지침에 따라 순수/보일러 배수 회수 제어 및 제약 용수 전도도 측정, 모니터링하는 데 적합 합니다.

- 측정 범위(전도도) : 0 ~ 2000μs/cm, 0 ~ 200.0ms/cm
                                  (TDS 환산 0 - 2000 mg/L)
- 온도 : 0 ~ 100℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Tough

  • Responds to harsh conditions at the site process
      Operating temperature range(Transmitter):-20 - 60℃
      Rated voltage: DC24V
      Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Sensor error can be detected immediately by self-diagnosis function
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS
  • Flexible temperature compensation

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring
Product name Conductivity meter
Model HE-300C
Sensors in combination 2-electrode method conductivity sensor
ESH and FS Series sensor of cell constant 0.01/cm, 0.1/cm, or 1.0/cm
Measurable
range
(*Note 1)
Cell constant (/cm) 0.01
(ESH-001)
0.1
(ESH/FS-01)
1.0
(ESH-1)
  Unit
Conductivity μS/cm 0.000 to 2.000
0.00 to 10.00
0.000 to 2.000
0.00 to 20.00
0.0 to 100.0
0.0 to 200.0
0 to 1000
mS/m 0.0000 to 0.2000
0.000 to 1.000
0.0000 to 0.2000
0.000 to 2.000
0.00 to 10.00
0.00 to 20.00
0.0 to 100.0
TDS
conversion
mg/L 0.00 to 2.00
0.0 to 10.0
0.00 to 2.00
0.0 to 20.0
0 to 100
0 to 200
0 to 1000
Temperature 0 to 100 (display range: −10℃ to 160℃)
Display
resolution
Conductivity
/ TDS conversion
As shown in the measuring table.
Temperature 0.01℃
Performance Conductivity Repeatability Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor
Linearity Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor
TDS conversion Repeatability Within ±1.5% of full-scale value (response for equivalent input)
Linearity Within ±1.5% of full-scale value (response for equivalent input)
Temperature Repeatability ±0.1℃(response for equivalent input)
Linearity ±0.5℃(response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC input/output insulation type (two-wire transmission type)
Load resistance Maximum: 600 Ω. Case of 24 V DC power supply (*Note 3)
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Conductivity: Free setting within a measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact
input
Number of input points 1
Contact type No-voltage a-contact
Conditions ON resistance: 40 Ω max.
Open voltage: 1.2 VDC
Short-circuit current: 21 mA DC max.
Contact function External input for transmission holding.
Temperature
compensation
Temperature element Platinum resistor: 1 kΩ (0℃)
Temperature compensation
  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95℃)
  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: ±3%/℃)
  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)
Temperature compensation
range
0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Conductivity Based on parameter input of coefficient for the sensor cell constant.
TDS conversion Conversion by user-defined coefficient value
(TDS calculation is 0.30 to 1.00 times of sample conductivity by μS/cm.)
Self-check Sensor diagnosis error Temperature sensor short-circuit, temperature sensor disconnection,
and out of the temperature measurement range
Meter error CPU error, ADC error, and memory error
Operating temperature range −20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature −25℃ to 65℃
Power
supply
Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 3)
Power consumption 0.6 W max.
Compatible
standards
CE marking EMC directive: EN61326-1
RoHS Directive: EN50581
  Immunity
Industrial
electromagnetic
environment
Electostatic discharge IEC61000-4-2
Electromagnetic field of radiated radio frequency IEC61000-4-3
Electric fast transient/burst IEC61000-4-4
Surge IEC61000-4-5
(*Note 2)
Conducted interference induced by radio frequency IEC61000-4-6
Emission
Class A
Radiated disturbance CICPR 11 CLASS A
Noise terminal voltage CISPR 11 CLASS A
FCC Rules Part15 CLASS A
Structure Installation Outdoor installation type
Installation method 50 Mount on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS 304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 2.8 kg; cover and mounting brackets: Approx. 1 kg

Note

  • Note 1: Only the sanitary sensor (FS-01 series) is applied.
  • Note 2: Conductivity measurement and TDS conversion measurement cannot be selected at the same time.
  • Note 3: The maximum load resistance that can be connected is decided depending on the powersupply voltage.

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • Note 4: When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 5: An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.