HP-300

현장 설치형 pH미터 (2-Wire Transmission)

HP-300은 pH센서를 연결하여 시료수의 pH와 온도를 측정하는 제품입니다.

다양한 생산 공정 및 폐수 처리 등의 모니터링 및 제어에 적합하며, 시료 특성에 맞게 다양한 유형의 전극을 사용할 수 있습니다.

원터치 기능을 장착해서 보정을 더욱 쉽게 지원하여 현장에서의 작업 시간을 대폭 단축시켰습니다.
또한, 보정 히스토리 확인이 가능합니다.

- 측정 범위

pH : 0 ~ 14
온도 : 0 ~ 100 ℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Tough

  • Responds to harsh conditions at the site process
      Operating temperature range(Transmitter):-20 - 60℃
      Rated voltage: DC 24
      Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Sensor error can be detected immediately by self-diagnosis function
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring
Product name pH meter
Model HP-300
Combined
sensor
Glass electrode
Measurable range pH pH0 to pH14 (display range: pH−1 to pH15)
Temperature 0℃ to 100℃
Display range when automatic detection capability for each temperature
sensor type is used: −10℃ to 110℃
Display range when temperature sensor type is manually set: −20℃ to 130℃
Display
resolution
pH 0.01 pH
Temperature 0.1℃
Performance pH Repeatability Within ±0.03 pH (response for equivalent input)
Linearity Within ±0.03 pH (response for equivalent input)
Temperature Repeatability ±0.3℃ (response for equivalent input)
Linearity ±0.3℃ (response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 600Ω Case of 24 V DC power supply (*Note 1)
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.08 mA (output only)
Output range pH: Selection from preset ranges or free range input within measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact input Number of input points 1
Contact type No-voltage "a" contact for open collector
Conditions ON resistance: 40 Ω 
Open voltage: 1.2 V
Short-circuit current: 21 mA DC max.
Contact function External input for transmission holding.
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature
:500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃)
Element selection method Automatic temperature sensor type detection or manual selection
(No temperature element is selectable)
Temperature compensation
range
0℃ to 100℃
Temperature calibration 1 point calibration comparing reference thermometer
Calibration Calibration method Auto or basic (manual) calibration
Number of calibration points Selectable from 1, 2, and 3 points
Kinds of standard solutions pH 2, 4, 7, 9, 10
Any pH standard solution (with difference of 2 pH or more) for basic calibration
Additional capabilities Automatic detection of kind of standard solution
Automatic detection of electric potential stability
Automatic detection of calibration failure
(asymmetry potential, sensitivity, or response time)
Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration)
Self-check Calibration error Asymmetry potential error, sensitivity error, and response time error
Temperature calibration error
Standard solution detection error
Electrode diagnostic error Cracking of glass response membrane
Reference electrode impedance error (only applicable for differential circuit mode with a liquid ground electrode)
Temperature sensor short-circuit, temperature sensor disconnection, and deviation from temperature measurement range
Meter error CPU, ADC, and EEPROM errors
Operating temperature range −20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature −25℃ to 65℃
Power supply Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 1)
Power consumption 0.6 W max.
Compatible
standards
CE marking Compatible
standards
EMC directive EN61326-1
RoHS Directive EN50581
Immunity Industrial electromagnetic environment
Electostatic discharge IEC61000-4-2
Electromagnetic field of radiated
radio frequency
IEC61000-4-3 (*Note 2)
Electric fast transient/burst IEC61000-4-4 (*Note 2)
Surge IEC61000-4-5 (*Note 3)
Conducted interference induced
by radio frequency
IEC61000-4-6 (*Note 2)
Emission Class A
Radiated disturbance CICPR 11 CLASS A
FCC rules Part15 CLASS A
Structure Installation Outdoor installation type
Installation method Mount on 50 A pole or wall
Protection code IP65 IEC60529, JIS C0920
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 2.8 kg
cover and mounting brackets: Approx. 1 kg

Note

  • Note 1:  The maximum load resistance is available within the range shown in the figure below, depending on power supply voltage:

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • Note 2:  The standard for effect on the reading by the electromagnetic field of the radiated radio frequency and by the conducted interference is within the measured pH value 0.1 pH.
  • Note 3:  When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 4:  An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.