Size:
0.37
MB
Brochure-SLIA-5000_HAER0266A
실리카 측정기
실리카(SiO2) 관리는 스케일 형성을 막는 데 가장 중요합니다. 터빈에 존재하는 규모가 특히 화력발전소와 원자력발전소의 발전 효율을 떨어뜨리기 때문입니다.
HORIBA의 다년 간의 트랙 기록과 기술적 노하우로 실리카 분석기는 기존의 안정성, 신뢰성 및 유지보수성을 보강하여 재탄생했습니다.
| General specifications | Equipment name | SILICA ANALYZER | |
|---|---|---|---|
| Model | SLIA-5000 | ||
| Dimensions | 600 (W) mm × 750 (H) mm × 1600 (D) mm | ||
| Mass | Approximately 130 kg | ||
| Power supply | 100 V AC ±10%, 50/60 Hz (Option) 110 V AC/115 V AC/120 V AC/220 V AC/230 V AC/240 V AC/ 240 V AC±10%, 50/60 Hz | ||
| Power consumption | Approximately 500 VA | ||
| Performance | Measurement range |
| |
| Resolution (Unit setting: μg/L) | 0.1 μg/L | Data<100 μg/L | |
| 1 μg/L | 100 μg/L≦Data≦5000 μg/L | ||
| Resolution (Unit setting: mg/L) | 0.0001 mg/L | Data<0.1 mg/L | |
| 0.001 mg/L | 0.1 mg/L≦Data≦5.0 mg/L | ||
| Repeatability | ±2% of the full scale | ||
| Measurement principle | Molybdenum blue method | ||
| Number of measurement points | 1 point to 4 points (specified when receiving the order) | ||
| Measurement time | 5 minutes | ||
| Measurement cycle | Specified as desired between 5 minutes and 999 minutes | ||
| Calibration | Zero | Each measurement | |
| Span | Equalizing filter method | ||
| Reagent replacement cycle | Once a week (reagent storage: 30℃ or below) | ||
| Installation conditions | Ambient temperature | 0℃ to 45℃ | |
| Ambient humidity | Relative humidity 85% or less (no condensation) | ||
| Ambient environment |
| ||
| Measurement conditions | Sample conditions | Temperature | 2℃ to 40℃ (36°F to 104°F) |
| Flow rate | 0.5 L/min to 1.0 L/min | ||
| Pressure | 0.02 MPa to 0.20 MPa | ||
| pH value | 11.0 or less | ||
| Coexisting phosphate ion concentration | 20 mg/L or less | ||
| However, no phosphate ions are permissible for the 0 μg/L to 10 μg/L/ 0 μg/L to 100 μg/L range (ultrapure water specification) | |||
| External input/output | Analog output | Concentration output | |
| 4 mA DC to 20 mA DC or 0 mA DC to 16 mA DC, selection setting in the operation section | |||
| Marker/hold output, selection setting in the operation section | |||
| Maximum load of 500 Ω or less, input/output isolation type, nonisolated between channels | |||
| Option: 0 V DC to 1 V DC, 1 V DC to 5 V DC | |||
| Contact output | Standard | a-contact output (b-contact can be specified when receiving the order) | |
| Contact rating | 250 V 3 A AC, 30 V 3 A DC | ||
| Measurement line shared | Meas., Standby, Maintenance, External control, Sync. Signal, Equipment Err., Power, etc. | ||
| Measurement line individual | Line, Sample Lack, Abnormal concentration, etc. | ||
| Contact input | No-voltage a-contact input (open collector is available) | ||
| Isolated input: Common to the (-) side | |||
| ON resistance | 100 Ω (max.) | ||
| Open-circuit voltage | 26 V DC (max.) | ||
| Short-circuit current | 13 mA DC (max.) | ||
| Measurement line shared | Meas. Start, Cal. Start | ||
| Measurement line individual | MEAS. LINE SET, MEAS. RANGE SET | ||
| Equipment function | Display | Color touch panel type liquid crystal | |
| Memory function | Measurement value | Memory for over 1 year | |
| Alarm history | 500 events | ||
| Calibration history | 100 events | ||
| Operation history | 100 events | ||
| USB external memory device can be used | |||
| Option | Filter for sample line (For Line 1 to Line 4) | By installing about 5 μm cartridge filter in the upstream of the overflow tank, it is possible to prevent iron rust, etc. from entering the equipment.
| |
| Flowmeter for sample line (For Line 1 to Line 4) | By adding this, you can accurately measure the sample supply flow rate. | ||
HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.
