Technical
Weight | 170 kg |
Power requirements | 100/240 V 50 Hz to 60 Hz - 800 VA |
Dimensions | 73 x 170 x 55 (H x W x D) cm |
Printer | Laser |
Throughput | 120 tests/hour |
Data processing | Color LCD 12.1 in |
Compliance | ASTM protocol |
Samples
Parameters (32) | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*, NEU, LYM, MON, EOS, BAS, ALY, LIC, IMM, IMG, IML: (% and #) |
Graphs | High-definition DIFF matrix, PLT, RBC, WBC/BAS curves |
Volume | Manual mode : 130 µL |
Stability | Up to 48 hours for CBC/DIFF |
Operating sample mode | Autoloader 15 racks (150 tubes) capacity |
Identification | Alphanumeric, Bar-code reader (internal and external) |
Results
Parameters | Presicion (CV %) | Linearity |
---|---|---|
WBC | < 2 | 0 - 150 103 / mm3 |
RBC | < 2 | 0 - 8 106 / mm3 |
HGB | < 1 | 0 - 24 g/dL |
HCT | < 2 | 0 - 67% |
PLT | < 5 | 0 - 1900 103 / mm3 |
PLT (PLT concentrate) | 0 - 2800 103 / mm3 | |
LYM | < 5 | -- |
MON | < 10 | -- |
NEU | < 3 | -- |
EOS | < 15 | -- |
BAS | < 20 | -- |