〈Catalog〉Industrial Water Quality Measuring Instruments
Field-installation Type ORP Meter (Two-Wire Transmission)
HO-300 connects ORP sensor and measures ORP and temperature in the sample water. It is suitable for neutralization reaction treatment management of aeration tank, denitrification tank, etc.
ORP: -2000 - 2000 mV
Temperature: 0 - 100℃
Product name | ORP meter | |||
---|---|---|---|---|
Model | HO-300 | |||
Combined sensor | ORP electrode | |||
Measurable range | ORP | −2000 mV to 2000 mV (display range: −2200 mV to 2200 mV) | ||
Temperature | 0℃ to 100℃
| |||
Display resolution | ORP | 1 mV | ||
Temperature | 0.1℃ | |||
Performance | ORP | Repeatability | Within ±5mV (response for equivalent input) | |
Linearity | Within ±5mV (response for equivalent input) | |||
Temperature | Repeatability | ±0.3℃ (response for equivalent input) | ||
Linearity | ±0.3℃ (response for equivalent input) | |||
Transmission output | Output type | 4 mA to 20 mA DC: input/output insulation type | ||
Load resistance | Max.: 600Ω (case of 24 V DC power supply)*1 | |||
Linearity | Within ±0.08 mA (output only) | |||
Repeatability | Within ±0.02 mA (output only) | |||
Output range | ORP: Selection from preset ranges or free range input within measuring range. | |||
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA | |||
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. | |||
Contact input | Number of input points | 1 | ||
Contact type | No-voltage a-contact | |||
Conditions | ON resistance: 40 Ω Open voltage: 1.2 V Short-circuit current: 21 mA DC max. | |||
Contact function | External input for transmission holding. | |||
Temperature measurement | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) | ||
Positive relation resistor with temperature: 500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃) | ||||
Element selection method | Automatic temperature sensor type detection or manual selection (no temperature element is selectable) | |||
Calibration | ORP adjust function | Adjust (offset adjustment): −200 mV to 200 mV Span sensitivity correction: 0.500 to 1.500 | ||
Temperature calibration | 1 point calibration comparing reference thermometer | |||
Self-check | Calibration error | Temperature calibration error Out of temperature calibration range (for ORP sensor with RTD). | ||
Electrode diagnostic error | Reference electrode impedance error (only applicable for differential circuit mode with a liquid contact electrode) Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range | |||
Converter error | CPU, ADC, and memory errors | |||
Operating temperature range | −20℃ to 60℃ (without freeze) | |||
Operating humidity range | Relative humidity: 5% to 90% (without condensation) | |||
Storage temperature | −25℃ to 65℃ | |||
Power supply | Rated voltage | 24 V DC (operating voltage range: 21 V to 32 V DC)*1 | ||
Power consumption | 0.6 W or less. | |||
Compatible standards | CE marking | |||
Compatible standards | EMC directive: EN61326-1 RoHS Directive: EN50581 | |||
Immunity | Industrial electromagnetic environment | |||
Electostatic discharge | IEC61000-4-2 | |||
Electromagnetic field of radiated radio frequency | IEC61000-4-3*2 | |||
Electric fast transient/burst | IEC61000-4-4 | |||
Surge | IEC61000-4-5*3 | |||
Conducted interference induced by radio frequency | IEC61000-4-6*2 | |||
Emission | Class A | |||
Radiated disturbance | CICPR 11 CLASS A | |||
FCC Rules | Part15 CLASS A | |||
Structure | Installation | Outdoor installation type | ||
Installation method | Mounted on 50 A pole or wall | |||
Protection code | IP65 | IEC60529, JIS C0920 | ||
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) | |||
Material of fittings | SUS304 | |||
Material of hood | SUS304 (coated with epoxy modified melanine resin) | |||
Material of window | Polycarbonate | |||
Display element | Reflective monochrome LCD | |||
External dimensions | 180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets) | |||
Mass | Main body: approx. 2.8 kg, Hood and mounting brackets: approx. 1 kg |
Though an arrester (spark over voltage: 400 V) is implemented for transmission output, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.
Sie haben Fragen oder Wünsche? Nutzen Sie dieses Formular, um mit unseren Spezialisten in Kontakt zu treten. * Diese Felder sind Pflichtfelder.