| PHYSICAL SPECIFICATIONS | |
| Dimensions & Weight | W*D*H (mm): 397(W) × 477(D) × 483(H) mm, 25.8kg |
| Printer (optional) | Compatible models with Linux drivers |
| Sample Management | Closed Tube/sample holder |
| Analysis Mode | CBC, DIFF CBC+CRP, DIFF+CRP, CRP |
| Sound Level | 57 dBa |
| Operating Temperature & Humidity | Operating Temperature & Humidity: +18°C (64+°F) to + 30°C (+86°F) Relative humidity ~80%, without condensation |
| Specimen Type & Volume | Human venous and capillary blood Whole blood, serum or plasma from vein for CRP measurement EDTA K2, EDTA K3 anticoagulant CBC, DIFF: 20 μL CBC+CRP, DIFF+CRP, CRP: 27 μL |
| Power Requirements | Power Supply: AC 100-240V (within ±10%), 50/60Hz Power consumption: 190 VA Heat output: 378 KJ/h |
| Reagents | ABX Diluent (10L or 20L), Whitediff 1L (cyanide free), ABX Cleaner 1L, CRP UNIT50 (50 tests * 2 pcs), 1 reagent for concentrated cleaning: ABX Minoclair |
| ECO FRIENDLY SOLUTION |
| LED light source Cyanide free lysis reagent WEEE Standard compliance |
| MEASUREMENT PRINCIPLES | |
| WBC & Differential | |
| Methods |
|
| HGB Measurement Method |
|
| RBC & PLT Detection Methods |
|
| HCT Measurement Method |
|
| CRP | |
| Latex immunity turbidity rate method | |
| PARAMETERS & PERFORMANCE DATA | |
38 Parameters WBC NEU# NEU% LYM# LYM% MON# MON% EOS# EOS% BAS# BAS% IMG# IMG% IML# IML% IMM# IMM% ALY# ALY% LIC# LIC% RBC HGB HCT MCV MCH MCHC RDW-CV RDW-SD MIC MAC PLT MPV PCT PDW P-LCC P-LCR CRP | |
| SOFTWARE SPECIFICATIONS | |
| Data Processing | Color LCD touch screen: 12.1 inches Operating System: Linuxâ„¢ Connection: RS232, Ethernet, USB Communication: ASTM & HL7 protocols Capacity: 10 000 results + graphs Options: keyboard, mouse and barcode reader |
| Quality Control | CBC/DIFF - 3 controls levels (Low, Normal, High) QC results compatible with HORIBA Quality Control Program (QCP), Levey-Jennings graphs, Radar graphs, XB on 3 or 9 parameters, Overlapping QC (DiFF - 6 active QC files) |
| ANALYTICAL MEASUREMENTS | |||
| Linearity | Linearity Limits | Visible Range | Unit |
| WBC | 0-300 | 300-999 | 109/L |
| RBC | 0-8 | 8-18 | 1012/L |
| HGB | 0.0-24.0 | 24.0-30.0 | g/dL |
| HCT | 0-67 | 67-80 | % |
| PLT (HGB >1.5 g/dL) | 0-2500 | 2500-4000 | 109/L |
| PLT (HGB <1.5 g/dL) | 0-4000 | 4000-5000 | 109/L |
| CRP (HCT ≥23%) | 0.0-200.0 | 200.0-230.0 | mg/L |
| CRP (HCT <23%) | 0.0-150.0 | 15.0-170.0 | mg/L |
| Limit of quantitation | Value | Unit | |
| WBC | 0.2 | 109/L | |
| RBC | 0.2 | 1012/L | |
| HGB | 1 | g/dL | |
| HCT | 2 | % | |
| PLT | 10 | 109/L | |
| CRP | 1 | mg/L | |
| Precision | CV% | in Range | Unit |
| WBC | 3 | 4-10 | 109/L |
| RBC | 2 | 3.6-6.2 | 1012/L |
| HGB | < 1.5 | 12.0-18.0 | g/dL |
| HCT | < 2.0 | 36-54 | % |
| PLT | < 5.0 | 180-500 | 109/L |
| CRP | < 10.0 | 0-10.0 | mg/L |
| CRP | < 8.0 | 10-120.0 | mg/L |