U-50

Multi-parameter water quality checker

The ability to simultaneously measure and display up to 11 parameters using a single device.

The Multi Checker U-50 series is a field-capable water quality checker that consolidates various sensors into a single probe, allowing for simultaneous measurement of up to 11 parameters in a single test. The design of the instrument takes into consideration ease of use in field measurements, and it is equipped with diverse functions and capabilities. It can be easily utilized by anyone for various applications, including river water, groundwater, and wastewater, where water quality measurement and testing are required.

pHORPCONDDOSALTTURBGPSSeawater Specific GravityTDSWater DepthWater Temperature

*

*Optional features

Data Collection Software

Please provide your own batteries. Batteries are not included with this product.

Division: Water Pollution
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Various sensors have been consolidated into a single probe

Simultaneous measurement and display of up to 11 parameters are possible. The sensors can be individually replaced.
However, only one type of standard solution is required.

Can be easily operated by anybody

Equipped with a one-touch automatic calibration function. The cable connection is an easy-to-attach/detach one-touch type. The data is conveniently displayed in real-time through the Auto-Hold function, allowing for easy visual monitoring during measurements. Furthermore, it is equipped with a diagnostic function that indicates the replacement timing of each sensor and displays errors.

Easy and convenient data management function

Easy computer integration is possible with USB connectivity.

The LCD display makes it easy to read and operate

The data is displayed clearly on the LCD screen for easy viewing. Possible to operate easily with one hand. Various functions are displayed with easily recognizable icons.

Convenient for field measurements

Designed for field measurements, it having been enhanced with impact resistance using soft resin materials. Also equipped with a convenient carrying bag as a standard accessory.

The turbidity sensor (U-53) is compliant with EPA Method 180.1

The U-53 incorporates a turbidity sensor that is compliant with EPA Method 180.1. Equipped with a wiper.

  U-51 U-52 U-52G U-53 U-53G
Sensor Probe Measurement temperature −10 to 55℃
Maximum sensor diameter Approx. 96 mm
Probe length Approx. 340 mm
Cable length Standard: 2 m, option: 10, 30 m
Mass Approx. 1,800 g (Approx. 3.97 lbs)
Automatic calibration (uses pH4) ●
Turbidity wiper — — — ● ●
Measurement depth Max. 30 m
Liquid contact part material (liquid end material) PPS, glass, SUS316L, SUS304, FKM, PEEK,Q, titanium, FEP membrane, POM
Water resistance JIS protection level 8
Control Unit Outer dimensions 115 (W) × 66 (D) × 283 (H) mm
Mass Approx. 800 g (approx. 1.76 lb)
LCD 320 × 240 liquid crystal display; backlit, black and white
Data memory 10,000
Communication USB
Battery C batteries x 4
Water resistance JIS protection level 7 (when sensor cable is fitted)
Battery Life Approx. 70 hours (without backlight) Approx. 500 measurements
Storage temperature −10 to 60℃
Ambient temperature −5 to 45℃
pH
●Two-point calibration
●Automatic temperature
Measurement principle Glass electrode method
Range pH0 to 14
Resolution 0.01 pH
Repeatability ±0.05 pH
Accuracy ±0.1 pH
Oxidation
Reduction Potential
(ORP)
Measurement principle Platinum electrode method
Range −2000 mV~+2000 mV
Resolution 1 mV
Repeatability ±5 mV
Accuracy ±15 mV
Dissolved Oxygen (DO)
●Salinity conversion
(0 to 70 PPT/automatic)
●Automatic temperature
compensation
Measurement principle Polarographic method
Range 0 to 50.0 mg/L
Resolution 0.01 mg/L
Repeatability ±0.1 mg/L
Accuracy 0 to 20 mg/L: ±0.2 mg/L
20 to 50 mg/L: ±0.5 mg/L
Conductivity (COND)
●Auto range
●Automatic temperature
conversion (25℃)
Measurement principle 4 AC electrode method
Range 0 to 10 S/m (0 to 100 mS/cm)
Resolution 0.000 to 0.999 mS/cm: 0.001 mS/cm
1.00 to 9.99 mS/cm: 0.01 mS/cm
10.0 to 99.9 mS/cm: 0.1 mS/cm
Repeatability 0.0 to 99.9 mS/m: 0.1 mS/m
0.100 to 0.999 S/m: 0.001 S/m
1.00 to 9.99 S/m : 0.01 S/m
Accuracy *±1% F.S. (Median of two-point calibration)
Salinity Measurement principle Conductivity conversion
Range 0 to 70 PPT (permillage)
Resolution 0.1 PPT
Repeatability ±1 PPT
Accuracy ±3 PPT
Total Dissolved Solid
(TDS)
●Conversion factor setting
Measurement principle Conductivity conversion
Range 0 to 100 g/L
Resolution 0.1 % F.S.
Repeatability ±2 g/L
Accuracy ±5 g/L
Seawater specific
gravity
●Display σt, σ0, σ15
Measurement principle Conductivity conversion
Range 0 to 50σt
Resolution 0.1 σt
Repeatability ±2 σt
Accuracy ±5 σt
Temperature Measurement principle Thermistor method
Range −10 to 55℃
Resolution 0.01 ℃
Repeatability *±0.10℃ (at calibration point)
Accuracy JIS class B platinum thermometer sensor (±0.3+0.005 | t |)
Turbidity (TURB) Measurement principle — LED light source and 30°scattering method Tungsten lamp source and 90°scattering method
Range 0 to 800 NTU 0 to 1000 NTU
Resolution 0 to 99.9 NTU: 0.1 NTU
100 to 800 NTU: 1 NTU
0 to 9.99 NTU: 0.01 NTU
10 to 99.9 NTU: 0.1 NTU
100 to 1000 NTU: 1 NTU
Repeatability *±5% (Reading) or ± 0.5 NTU whichever is greater ±3% (Reading) or ±0.1 NTU whichever is greater
Accuracy ±5% (Reading) or ±1 NTU whichever is greater 0 to 10 NTU: ±0.5 NTU
10 to 1000 NTU: 3% (Reading)
or ±1 NTU whichever is greater
Water depth Measurement principle — — Pressure method
Range 0 to 30 m
Resolution 0.05 m
Repeatability ±1 % F.S.
Accuracy ±0.3 m
GPS 12 channel parallel — — ● — ●

Note:
* Battery life based on continuous operation using alkaline C dry batteries when the monitor temperature is over 20℃ and the backlight OFF.
* Accuracy is measured by calibrating 4 points for turbidity and electrical conductivity and 2 points for all other measurements against standard solution.
* Repeatability is measured by the ability to reproduce the results against the standard solution (at 25℃ normal pressure condition).