Pentra MS CRP

HORIBA Medical’s CRP hematology combined solutions at Point of Care Testing

Hematology Analyzer

How will you contribute to the reduction of antibiotic resistance?

  • 5 DIFF + CRP on the same platform
  • Micro-sampling on whole blood: 5 DIFF mode : 26μL / 5 DIFF +CRP: 35μL
  • Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology.
  • Perfect homogenization of blood samples with reagents - MDSS** Technology.
  • CRP:Latex Absorbance measurement  at  660nm Wavelength with Hematocrit correction

 

* DHSS : Double Hydrodynamic Sequential System
** MDSS : Multi-Distribution Sampling System 

*This product may not be available in your country or region at this time. Please contact your HORIBA Medical sales representative or distributor for more information. 

Segment: Medical
Division: Hematology
Manufacturing Company: HORIBA, Ltd.
  • Throughput: Up to 60 samples/hour
  • Reagents: Only 4 onboard reagents and 1 diluent
  • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
  • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
  • Basophils counted through specific channel
  • High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)    

 

* DHSS : Double Hydrodynamic Sequential System
*** RUO parameters (Research Use Only)

 

 

Technical

Weight

38 kg

Power supply

100 to 240 V (± 10%) - 50 to 60 Hz.

Dimensions

51,6 cm x 44,5 cm x 58 cm

Printer

General-purpose printer used with Windows PC

Power Consumption:

160W (200VA)

Interface:

USB connector (for PC connection)

Throughput

DIFF mode, QC (DIFF) mode: approx. 60 samples/hour

DIFF+CRP mode : approx. 20 samples/hour

Data management

Color screen: 15 in. monitor (800 x 600px)

Windows XP Prof.TM 

User-defined flagging limits

Transmit patient & QC to LIS

Uni-directional or bi-directional connection

Communication Protocols: ABX / ASTM

CD-ROM drive / Floppy disk

2 serial ports minimum

USB Keyboard / USB Mouse

Memory

Result: 1000 results x 500 files

Calibration result: 11 results x 6 lots x 2 types

QC result: 400 results x 6 lots x 4 types

XB result: 20 results x 60 batches

Quality Control Management

12 selectable QC files

XB: 60 Operator selectable batches with statistics (20 samples per batch)

Within run

Levey-Jennings Graphs

Logs

Reagents, calibration, maintenance, errors, blank cycle

 

 Parameters (27)

CRP, WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*.

Graphs

High-definition LMNE matrix
PLT, RBC, WBC/BAS curves.

Blood Volume

CBC + DIFF mode: 26 µL
CBC + DIFF + CRP mode : 35 µL

Stability

48 hours post-draw stability

Operating sample mode

Closed tube

Identification

Alphanumeric. Bar-code reader (optional)

 * RUO parameters (Research Use Only)

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Product accessories

Accreditation Folder (ISO)
Accreditation Folder (ISO)

Accreditation Support

CellaVision®
CellaVision®

Digital Cell Morphology Systems

Hematovision
Hematovision

Accreditation Support

Quality Control Program (QCP)
Quality Control Program (QCP)

Accreditation Support

Corporate