Synapse EM Camera

EMCCD for low-light and ultrafast spectroscopy

UV-VIS-NIR EMCCDs Camera

EMCCD Camera for OEM Industrial Applications

Key Applications

  • Raman spectroscopy
  • SERS, TERS
  • Bioluminescence
  • Single-molecule spectroscopy
Π‘Π΅Π³ΠΌΠ΅Π½Ρ‚: Scientific
ΠŸΠΎΠ΄Ρ€Π°Π·Π΄Π΅Π»Π΅Π½ΠΈΠ΅: Mini-Spectrometers and Scientific Cameras for OEMs
ΠŸΡ€ΠΎΠΈΠ·Π²ΠΎΠ΄ΠΈΡ‚Π΅Π»ΡŒ: HORIBA Scientific

Key Features and Benefits

  • 1600 Γ— 200 or 1600 Γ— 400 EMCCD sensor  - 16 Γ— 16 ΞΌm pixel size for high spectral resolution
  • 25.6 Γ— 3.2 mm or 25.6 Γ— 6.4 mm image area  - Ideal for high-speed or multi-track spectroscopy
  • Back- and front-illuminated - Back-illuminated for high QE or front-illuminated for etalon fringe suppression. 
  • Deep thermoelectric cooling  - Air or liquid circulation to minimize dark current
  • Dual readout modes  - EMCCD or CCD for a broad range of light conditions
  • Readout rates up to 3 MHz  - Acquires more than 1600 spectra per second
  • Single fused-silica vacuum window  - Minimizes reflection losses from UV to near-IR

 

 

SensorFront-illuminated (FI), back-illuminated (BI), scientific grade 1 
Active pixels1600 Γ— 2001600 Γ— 400
Pixel size16 ΞΌm Γ— 16 ΞΌm
Image area25.6 mm Γ— 3.2 mm25.6 mm Γ— 6.4 mm
Output node well capacity
      High Sensitivity mode
      Electron Multiplying mode
300,000 eβ€”
1,300,000 eβ€”
Register Well Depth
      High Sensitivity mode
      Electron Multiplying mode
450,000 eβ€”
800,000 eβ€”
Non-linearity
(measured at all speeds per camera)
< 0.75%
Readout Noise (eβ€”):  Typ. (Max)
      High Sensitivity mode:        EM off
      Electron Multiplying mode: EM off
      Electron Multiplying mode: EM on
50 kHz                    1 MHz                    3 MHz
2.7 (5)                     6.1 (9)                    8.5 (12)
8.0 (15)                     23 (35)                    38 (50)
< 1                         < 1                        < 1
Dark Current at β€”60Β°C (eβ€”/pixel/s)
      FI
      BI
< 0.002
< 0.01
Maximum spectra per second
      Full vertical bin
      ROI mode 20 rows
      ROI mode 8 rows
616
1475
1613
376
1475
1613
Software-adjustable gain (eβ€”/count)
      High Sensitivity mode
      Electron Multiplying mode
Selectable from 0.6 to 4.0
Selectable from 3.6 to 25
Electron multiplier gain1 to 1000, software-controlled
Digitization16-bit ADC
Vertical shift rates4.9, 9.6, 19 software-selectable(4)
Cooling at +20Β°C
      Air-cooled
      Liquid-cooling recirculator
β€”60Β°C (guaranteed)
β€”75Β°C (typical)
Power requirements
      AC-DC power supply (provided)
AC input 90–264 VAC, 47–63 Hz
DC output +9 V, 6.44 A maximum

 

(1) Region Of Interest (ROI) mode 8 rows
(2) Measured at -60Β°C
(3) Measured at all read-out speeds for each camera
(4) Some decrease in CTE may be observed at faster speeds.

 

 

Π—Π°ΠΏΡ€ΠΎΡΠΈΡ‚ΡŒ ΠΈΠ½Ρ„ΠΎΡ€ΠΌΠ°Ρ†ΠΈΡŽ

Π£ вас Π΅ΡΡ‚ΡŒ вопросы ΠΈΠ»ΠΈ поТСлания? Π˜ΡΠΏΠΎΠ»ΡŒΠ·ΡƒΠΉΡ‚Π΅ эту Ρ„ΠΎΡ€ΠΌΡƒ, Ρ‡Ρ‚ΠΎΠ±Ρ‹ ΡΠ²ΡΠ·Π°Ρ‚ΡŒΡΡ с нашими спСциалистами.

* Π­Ρ‚ΠΈ поля ΠΎΠ±ΡΠ·Π°Ρ‚Π΅Π»ΡŒΠ½Ρ‹ для заполнСния.

Related products

CoolOne
CoolOne

Dual Mode Analog/Photon Counting PMT

FHR Series
FHR Series

High Resolution Monochromators

iHR Series
iHR Series

Mid-Focal Length Imaging Spectrometers

MicroHR Series
MicroHR Series

Short Focal Length Imaging Spectrometers

Scientific Cameras
Scientific Cameras

Deep Cooled UV/Vis/NIR

Single Channel Detectors
Single Channel Detectors

Large choice of PMTs, solid state, photoelectric detectors for custom spectroscopy solutions

Synapse EMCCD Camera
Synapse EMCCD Camera

EMCCD Scientific Camera

Synapse InGaAs/Symphony II InGaAs
Synapse InGaAs/Symphony II InGaAs

Deep Cooled NIR Scientific Cameras

Syncerity VUV
Syncerity VUV

Deep Cooled Vacuum Ultra Violet Scientific Cameras

ZNV-7
ZNV-7

Π“Π΅Π½Π΅Ρ€Π°Ρ‚ΠΎΡ€ Π½ΡƒΠ»Π΅Π²ΠΎΠ³ΠΎ Π³Π°Π·Π°

Corporate