旨在以非接觸方式測量半導體製造過程中使用的化學物質,例如清潔和蝕刻。
透過在不直接接觸的情況下測量管道內流動的化學物質,實現安全穩定的濃度監測,避免化學洩漏或污染的風險。
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
聯絡我們
找到我們
Non-contact Measurement Structure
By preventing chemical leaks and contamination, safe and stable measurements become possible. Furthermore, installation can be performed without cutting the piping, reducing installation time and minimizing the workload during implementation.
Stable Measurement of High-temperature Chemicals
With a revamped optical system and improved computational processing, high-temperature chemicals (20 to 80°C) essential for cutting-edge chemical processing can be reliably measured with stability.
Prolonged Lifespan of Light Sources
Prolonged lifespan through the adoption of LED lamps (with a 5-year warranty) has been achieved.
Space-saving
Light sources and spectrometers are integrated into the cell unit, contributing to a space-saving design. The display unit is a panel-mount type.
| Model name | Non-contact type chemical concentration monitor | ||||
|---|---|---|---|---|---|
| Model | CS-900 | ||||
| Measurement method | Absorption spectroscopy | ||||
| Concentration calculation | Temperature compensated multivariate analysis | ||||
| Measurement range/ Repeatability (mass%) | Model name | Component | Measurable range | Repeatability | Analog output range |
| CS-900-1P-1P25P5 | NH3 | 0.00%-1.20% | ±0.05% | 0.00-10.00 | |
| H2O2 | 0.00%-5.50% | ±0.10% | 0.00-10.00 | ||
| H2O | 93.3%-100.0% | ±1.0% | - | ||
CS-900-1P-P8P8 | NH3 | 0.20%-0.80% | ±0.03% | 0.00-1.00 | |
| H2O2 | 0.20%-0.80% | ±0.03% | 0.00-1.00 | ||
| H2O | 98.4%-99.6% | ±0.3% | - | ||
Conditions of measurement | 1) Measurement Frequency: every 3 seconds 2) Moving average : 16 times | ||||
Tube | PFA tube 1inch or 3/4inch (Do NOT use anti-static type PFA tube) | ||||
Sample condition*1 | Sample temperture : 20 to 80°C Sample temperture, Atmospheric temperture fluctuation : ±1 degree C (period of time: 1 hour) Sample pressure : Less than 0.2 MPa (G) Sample pressure fluctuation : Within 0.02 MPa | ||||
Power source | DC 24V | ||||
Power consumption | Apporox. 30W (Transient electric current at the time of start is excluded) | ||||
| Weight | Display unit : Apporox. 0.9kg Sensor unit: Apporox. 2.9kg | ||||
| Dimension (WxDxH) | Display unit : 96x135x96mm Sensor unit: 238x135x190mm | ||||
| Slanting angle | Sensor unit: Vertical direction within ±3 degree Lateral direction 0 to 90 degree | ||||
| Light source | 5years | ||||
| communication | Parallel I/O, RS-232C, Analog output | ||||
如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。
For your specific application
Benchtop pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter