HD-200

Field-installation Type Dissolved Oxygen Meter (Four-Wire Transmission, Membrane Type)

HD-200 connects the DO (Dissolved Oxygen) sensor (DO-1100) and measures the DO in the sample water.The measurement method is membrane type polarography method. It is suitable for monitoring DO in wastewater treatment or aquarium for aquaculture.

Measurement range

Dissolved oxygen concentration: 0 - 20 mg/L
Saturation degree: 0 - 200 %
Temperature: 0 - 50 ℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Tough

  • Responds to harsh conditions at the site process
      Operating temperature range(Transmitter):-20 - 55℃
      Rated power supply voltage:AC 90 - 264 V ±10% 50/60 Hz
      Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Self-diagnosis of membrane damage and infiltration into the probe
  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring
  • On-site membrane replacement as well as whole sensor replacement
  • Span calibration by air preceded by auto-zeroing (Electrical auto zeroing function)
Product nameDissolved oxygen converter
ModelHD-200
Combined sensor5505, 5510 (A bipolar diaphragm type polarography sensor)
Combined probeDO-1100
Measurement
range
Dissolved oxygen
concentration
0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L)
Saturation degree0% to 200% (Display range: 0% to 200%)
Temperature0℃ to 50℃ (Display range: -10℃ to 110℃)
Display
resolution
Dissolved oxygen
concentration
0.01 mg/L
Saturation degree1%
Temperature0.1℃
PerformanceDissolved
oxygen
concentration
RepeatabilityWithin ±0.5% of the full scale (response for equivalent input)
LinearityWithin ±0.5% of the full scale (response for equivalent input)
TemperatureRepeatability±0.5℃(response for equivalent input)
Linearity±0.5℃(response for equivalent input)
Transmission
output
Number of output2 (The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type4 mA to 20 mA DC: input/output isolated type
Load resistanceMaximum: 900Ω
LinearityWithin ±0.08 mA (output only)
RepeatabilityWithin ±0.02 mA (output only)
Output rangeOutput 1Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range.
Output 2Free setting within a range between -20℃ and 130℃
Occasional out for errorHold or burnout to either 3.8 mA or 21 mA
Transmission holdIn the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact outputNumber of output3
Output typeNo-voltage contact output
Contact typeRelay contact; SPDT (1c)
Output capacity250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact
function
R1 and R2Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding of transmission output, and cleaning output (The contact is closed during alarm operation, opened normally and while the power is down).
FAILError alarm (Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting rangeSelectable from dissolved oxygen concentration, and temperature
  • Setting range: Within the measurement range
  • Delay time: 0 to 600 seconds
Control setting range
  • Setting range: 0 mg/L to 20 mg/L
  • Control width: 0.02 mg/L to 200 mg/L(±0.01 to ±0.50 mg/L)
Cleaning
output
Number of output points1
Output typeContact output with voltage (output of connected power supply voltage)
Contact typeRelay contact; SPST (1a)
Output contact capacity250 V AC 0.5 A
External instrumentSolenoid valve for cleaning control
SettingsCleaning period0.1 hours to 168.0 hours
Cleaning time2 seconds to 600 seconds
Hold time2 seconds to 600 seconds
Timer accuracyWithin 2 minutes per month
Description of cleaning
operation
One of the following operations.
  • Operation of internal timer only
  • Operation of internal timer and external input command
  • Enable the internal timer only when external input is active.
  • The cleaning trigger signal (The external trigger signal must be longer than 2 seconds)
Contact inputNumber of input1
Contact typeNo-voltage “a” contact of open collector
ConditionsON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact functionExternal input for cleaning or transmission holding if cleaner is not attached.
Transmission
capability
Communication typeRS-485
Signal type2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
(The temperature sensor is built into the probe DO-1000.)
Temperature compensation
range
0℃ to 50℃
Temperature calibration1 point calibration comparing to reference thermometer
CalibrationCalibration methodSpan calibration in the air or DO saturated aquaous solution
Number of calibration pointsAtmospheric calibration: 1 point
(Zero electric calibration is carried out automatically)
Saturation liquid calibration : 2 points (zero calibration is omitted)
Additional capabilitiesSalinity compensation (0.0% to 5.0%)
Automatic detection of calibration failure (Zero error, Sensitivity error)
Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity)
Self-checkCalibration errorZero calibration error, sensitivity error, outside of temperature calibration range
Electrode diagnostic errorSensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor
meter errorCPU error, CPU error, ADC error, memory error
Operating temperature range-20℃ to 55℃ (without freeze)
Operating humidity rangeRelative humidity: 5% to 90% (without condensation)
Storage temperature-25℃ to 65℃
Power supplyRated power supply voltage100 V to 240 V AC ±10% 50/60 Hz
Power consumption15 VA (max)
OthersWith power switch for maintenance use
Compatible
standards
CE markingEMC  :EN61326-1
      Class A, Industrial electromagnetic environment
Safety  :EN61010-1
RoHS :EN50581
      9. Industrial monitoring and control instruments
FCC rulesPart15 Class A
StructureInstallationOutdoor installation type
Installation methodMounted on 50A pole or wall
Protection codeIP65IEC60529, JIS C0920
Case materialAluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittingsSUS304
Material of hoodSUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of windowPolycarbonate
Display elementReflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg

*1:   The standard for effect on the reading by the electromagnetic field of the radiated radio frequency and by the conducted interference is within the measured dissolved oxygen value ±0.4 mg/L.

*2:   When the sensor cable, the transmission cable, or the contact input cable is extended to 30 m or longer, the surge test specified in the EMC directive for CE marking is not applied.

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.

Request for Information

Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.

Related products

CS-600F
CS-600F

Fiber Optic Type Chemical Concentration Monitor

CS-620F
CS-620F

Optical Fiber Type Hot Phosphoric Acid Concentration Monitor

CS-900
CS-900

Non-Contact Chemical Concentration Monitor

EMGA-921
EMGA-921

Hydrogen Analyzer

EMGA-Expert
EMGA-Expert

Oxygen/Nitrogen/Hydrogen Analyzers
(Flagship High-Accuracy Model)

EMGA-Pro
EMGA-Pro

Oxygen/Nitrogen/Hydrogen Analyzers (Entry Model)

HC-200F
HC-200F

Field-installation Type Simplified Fluoride Ion Concentration Meter(Four-Wire Transmission)

HC-200NH
HC-200NH

Field-installation Type Ammonia Nitrogen Meter(Four-Wire Transmission)

HC-300F
HC-300F

Field-installation Type Simplified Fluoride Ion Concentration Meter(Two-Wire Transmission)

HD-200FL
HD-200FL

Field-installation Type Dissolved Oxygen Meter (Four-Wire Transmission, Optical Type)

HD-300
HD-300

Field-installation Type Dissolved Oxygen Meter(Two-Wire Transmission, Membrane Type)

HD-480
HD-480

Panel-mount Type Dissolved Oxygen Meter (Four-Wire Transmission, Membrane Type)

HD-960L
HD-960L

Dissolved Oxygen Monitor in Low Concentration

HD-960LR
HD-960LR

Dissolved oxygen concentration monitor series for semiconductor manufacturing

HE-200C
HE-200C

Field-installation Type Electric Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-200H
HE-200H

Field-installation Type Electric Conductivity Meter (Four-Wire Transmission, For High-concentration)

HE-200R
HE-200R

Field-installation Type Electric Resistivity Meter (Four-Wire Transmission)

HE-300C
HE-300C

Field-installation Type Electric Conductivity Meter (Two-Wire Transmission, For Low-concentration)

HE-300R
HE-300R

Field-installation Type Electric Resistivity Meter (Two-Wire Transmission)

HE-480C
HE-480C

Panel-mount Type Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-480C-GC
HE-480C-GC

Carbon Sensor Conductivity Meter (Low concentration type)

HE-480H
HE-480H

Panel-mount Type Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-480R
HE-480R

Panel-mount Type Resistivity Meter(Four-Wire Transmission)

HE-480R(W)
HE-480R(W)

Industrial resistivity meter

HE-960CA
HE-960CA

Citric Acid Monitor

HE-960CW
HE-960CW

Panel-mount Type Conductivity Meter (Four-Wire Transmission, 2-channel)

HD-300
HD-300

Field-installation Type Dissolved Oxygen Meter(Two-Wire Transmission, Membrane Type)

HD-480
HD-480

Panel-mount Type Dissolved Oxygen Meter (Four-Wire Transmission, Membrane Type)

HD-960L
HD-960L

Dissolved Oxygen Monitor in Low Concentration

HD-960LR
HD-960LR

Dissolved oxygen concentration monitor series for semiconductor manufacturing

HE-200C
HE-200C

Field-installation Type Electric Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-200H
HE-200H

Field-installation Type Electric Conductivity Meter (Four-Wire Transmission, For High-concentration)

HE-200R
HE-200R

Field-installation Type Electric Resistivity Meter (Four-Wire Transmission)

HE-300C
HE-300C

Field-installation Type Electric Conductivity Meter (Two-Wire Transmission, For Low-concentration)

HE-300R
HE-300R

Field-installation Type Electric Resistivity Meter (Two-Wire Transmission)

HE-480C
HE-480C

Panel-mount Type Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-480C-GC
HE-480C-GC

Carbon Sensor Conductivity Meter (Low concentration type)

HE-480H
HE-480H

Panel-mount Type Conductivity Meter (Four-Wire Transmission, For Low-concentration)

HE-480R
HE-480R

Panel-mount Type Resistivity Meter(Four-Wire Transmission)

HE-480R(W)
HE-480R(W)

Industrial resistivity meter

HE-960CA
HE-960CA

Citric Acid Monitor

HE-960CW
HE-960CW

Panel-mount Type Conductivity Meter (Four-Wire Transmission, 2-channel)

HE-960H-KOH
HE-960H-KOH

KOH Monitor

HE-960H-TM
HE-960H-TM

Wide range TMAH Concentration Monitor

HE-960H-TM-S
HE-960H-TM-S

High Precision TMAH Concentration Monitor

HE-960HC
HE-960HC

Carbon Sensor Conductivity Meter (High concentration type)

HE-960HI
HE-960HI

Panel-mount Type Conductivity Meter (Four-Wire Transmission, For Wide Range)

HE-960LC
HE-960LC

Carbon Sensor Conductivity Meter

HE-960LF / FS-09F-1/2
HE-960LF / FS-09F-1/2

Flat Carbon Sensor Conductivity meter

HE-960R-GC(W)
HE-960R-GC(W)

Carbon Sensor Resistivity Meter

HE-960RW
HE-960RW

Panel-mount Type Resistivity Meter(Four-Wire Transmission, 2-channel)

HE-960RW-GC(W)
HE-960RW-GC(W)

2-Channel Resistivity Meter

HF-960H
HF-960H

HF / HCl Concentration Monitor

HF-960M
HF-960M

Low Concentration Type HF/HCl/NH3 Concentration Monitor

HF-SP-CU
HF-SP-CU

2-channel High Precision HF Concentration Monitor

HO-200
HO-200

Field-installation Type ORP Meter (Four-Wire Transmission)

HO-300
HO-300

Field-installation Type ORP Meter (Two-Wire Transmission)

HO-480
HO-480

Panel-mount Type ORP Meter (Four-Wire Transmission)

HR-200
HR-200

Feild-installation Type Residual Chlorine Meter (Four-Wire Transmission, Water flow beads)

HR-200RT
HR-200RT

Field-installation Type Residual Chlorine Meter (Four-Wire Transmission, Rotating Type)

HR-480
HR-480

Panel-mount Type Residual Chlorine Meter (Four-Wire Transmission)

HT-110
HT-110

Total Organic Carbon Monitor

HZ-960HPO-M
HZ-960HPO-M

H2O2 Monitor

LAQUA DS-72
LAQUA DS-72

Benchtop Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA EC1100
LAQUA EC1100

Benchtop Water Quality Meters

LAQUA F-74
LAQUA F-74

Benchtop pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUAtwin EC-11
LAQUAtwin EC-11

Pocket Water Quality Meters

LAQUAtwin EC-22
LAQUAtwin EC-22

Pocket Water Quality Meters

LAQUAtwin EC-33
LAQUAtwin EC-33

Pocket Water Quality Meters

Corporate