The MICROPOLE system is one of the smallest mass spectrometer systems making it easy to install. The system is proven to be an ideal solution for many chamber characterization applications. The compact size is achieved by the patented design of miniature array of quadrupole mass filters. The system is capable of operating at much higher pressures than traditional systems, so the need for additional differential pumping equipment is reduced.
Dimensions (mm)
![]()
Specifications
Micropole System
Spectrum Generator
| Model | QL-SG02-065-1A | QL-SG02-100-1A | QL-SG02-300-1A |
| Mass Range | m/z 2-65 | m/z 2-100 | m/z 4-300 |
| Mass Filter Type | Qadrupole | ||
| Detector | Faraday cup | ||
| Measurement Scan Speed | 50 - 6400 ms/m/z | ||
| Ionization Voltage | 43eV or 70eV selectable | ||
| Operating Temperature | 15 - 45°C | ||
| Operating Humidity | 80%RH or less (not condensing) | ||
| Storage Temperature | 0 - 80°C | ||
| Weight (with Special Clamp for SMPA Type Sensor) | 575g | ||
| Demension | W150 x H62 x D747 (mm) | ||
| RS485 communication baud rate | 115.2K / 38.4K / 9.6Kbps | ||
| Analog I/O interface | Input: 0-10V 2ch, Output: 0-10V 4ch | ||
| Contact I/O | Tool status 3ch, Control signal input 3ch, Set point output 5ch | ||
| Power input | DC 24V +/-5%, 100mVpp, 50W | ||
| Connector type | Round type connector: PRC03-23A10-4M | ||
* The below sensor specifications are based on a scanning speed of 400ms m/z unit.
* Windows 10 and higher based configuration and data acquisition software is included.
Sensor
| Model | MPA8-7-2/65C | MPA8-5-2/100C | SMPA8-1-2/100CR | MPA8-1-4/300C |
| Measured Upper Limit Pressue (N2)*1 | 0.5Pa | 0.2Pa | 0.1Pa | 0.2Pa |
| Maximum Operating Pressure (N2)*2 | 0.9Pa | 0.6Pa | 0.1Pa | 0.4Pa |
| Minimum Detectable Partial Pressure (N2) | 5.0×10-6Pa (70eV) | |||
| Resolution (FWHM: N2) | 1.2±0.3 m/z units | 1.0±0.3 m/z units | 1.5±0.3 m/z units | 1.8±0.3 m/z units |
| Maximum Bake out Temperature ( SG removed ) | 350°C | |||
| Maximum Operating Temperature ( SG installed ) | 150°C | |||
| Mounting Flange | ISO-KF16 (NW 16), CF34 (ICF 34) | |||
| Weight | MPA:CF34 Flange 50g
| SMPA:ISO-KF16 Flange 50g | MPA:CF34 Flange 50g | |
| Filament | Y2O3/Ir 2pcs | 3%Re/W 2pcs | Y2O3/Ir 2pcs | |
| RoHS | Compliance | |||
* SMPA type sensor is protected by a mesh on the vacuum fitting, designed to extend the lifetime of sensor in harsh environment such as plasma processes.
*1 A pressure at which the Q-mass signal output is saturated. Beyond this pressure, signal outputs may not correspond to actual pressures.
*2 An upper limit pressure that can be operated. If an operating pressure exceed the limit one, the measurement may stop to protect the sensor.