With the Ecodesign Regulations that came into effect on the 1st January 2022 for solid fuel room heater appliances, manufacturers of all solid fuel burning appliances will have to have their products independently tested to prove that they are compliant.
HORIBA enables manufacturers to carry out their own ‘in-house’ testing regimes to ensure every one of your products is within the new limits of the four main emission limit provisions as laid down within the new Ecodesign regulations. In-house testing allows you to have confidence that your appliances will pass the independent testing, therefore minimizing the costs from failed/repeat tests.
Introducing the VA-5000 Series Gas Analyzer with its key features:
*Refer to specifications table for possible combination of modules.
Measurement Method | Component | Option Range | Standard Range | Zero Drift | Span Drift*1 | Repeatability | |||
High Sensitive Min. Range | Min. Range | Max. Range | Standard Range | High Sensitivite | Standard Range | High Sensitive | |||
NDIR | CO | 0-50 ppm | 0-200 ppm | 0-100 vol% | ± 2.0%/week of F.S. | ± 2.0%/day (CO: 0-50-99 ppm range, CO2: 0-50-99 ppm range, SO2: 0-100-199 ppm range ± 1.0%/day (CO: 0-100-199 ppm range, CH4: 0-100-199 ppm range | ± 2.0%/week of F.S. | ± 2.0%/day (CO: 0-50-99 ppm range, CO2: 0-50-99 ppm range, SO2: 0-100-199 ppm range ± 1.0%/day (CO: 0-100-199 ppm range, CH4: 0-100-199 ppm range | ± 0.5% of F.S. |
CO2 | 0-50 ppm | 0-100 ppm | 0-100 vol% | ||||||
CH4 | 0-100 ppm | 0-200 ppm | 0-100 vol% | ||||||
N2O | NA | 0-100 ppm | 0-5000 ppm | ||||||
NO | NA | 0-500 ppm | 0-1 vol% | ||||||
SO2 | 0-100 ppm | 0-200 ppm | 0-10 vol% | ||||||
NH3 | NA | 0-100 ppm | 0-1000 ppm | ||||||
CLA | NO/NOx | NA | 0-20 ppm | 0-5000 ppm | ± 2.0%/week of F.S. | ± 0.5% of F.S. (Range is more than 0 ppm to 100 ppm) ± 1.0% of F.S. (Range is less than 0 ppm to 100 ppm) | |||
MPA | O2 | NA | 0-5 vol% | 0-100 vol% | ± 2.0%/week of F.S. | ± 2.0%/week of F.S. | ± 0.5% of F.S. | ||
Galvanic | NA | 0-5 vol% | 0-25 vol% | ± 1.0%/day of F.S. | ± 1.0%/day of F.S. | ± 0.5% of F.S. | |||
Zirconia | NA | 0-5 vol% | 0-25 vol% | ± 1.0%/week of F.S. | ± 1.0%/week of F.S. | ± 0.5% of F.S. | |||
PMA | NA | 0-5 vol% | 0-100 vol% | ± 2.0%/week of F.S. | ± 2.0%/week of F.S. | ± 0.5% of F.S. |
Note 1: Select multiple measurement ranges within the above minimum and maximum range table in accordance to the following conditions.
[NDIR] Five (5) ranges; the highest range must be within the maximum limit ratio of 10x the lowest range. Maximum limit of 20x the lowest range is also an available option, which may be limited by the cell length.
[CLA] Eight (8) ranges; the highest range must be within the maximum limit ratio of 100x the lowest range. If the maximum range exceeds 2000 ppm, the minimum range should be at least 50 ppm or more.
[MPA] Five (5) ranges; the highest range must be within the maximum limit ratio of 10x the lowest range.
[Galvanic] Five (5) ranges; the highest range must be within the maximum limit ratio of 5x the lowest range.
[Zirconia] Five (5) ranges; the highest range must be within the maximum limit ratio of 5x the lowest range.
[PMA] Three (3) ranges; the highest range must be within the maximum limit ratio of 10x the lowest range.
Note 2: Contact HORIBA if you require measurement of special gases or ranges.
*1% of span drift for NDIR is achievable with special adjustment at factory. Please contact the HORIBA team for further details.
* Please consult HORIBA for further details.
Mehrkomponenten-Gasanalysator für den Prozess und CEMS
Rauchgas Analysen System
Heated Type THC Analyzer