Hydrogen Gas Analyzer TCA-5000 | |
Measurement principle | Thermal conductivity detection method |
Measurement range | (Min.) 0 - 10 vol% |
Measurement range (optional) | 0 - 5 vol% to less than 10 vol% |
Repeatability | ± 1.0% of full scale |
Linearity | ± 1.0% of full scale |
Drift (zero/span) | ± 2.0% of full scale per week |
Response time (T90) | < 20 s from sample inlet |
Sample gas flow rate | Approx. 0.5 L/min |
Operation temperature | 0 to 40°C (32 to 104°F) |
Power supply | AC 100 V - 240 V ±10% (Max. voltage: AC 250 V) |
Analog output (optional) | 0 - 16 mA / 4 - 20 mA / 0 - 20 mA or 0 - 1 V |
* Interference on H₂ concentration by other coexisting in sample gas components can be corrected combining TCA-5000 with the Multi-Component Gas Analyzer VA-5000 series. Please consult with us.