HD-200

Field-installation type dissolved oxygen meter (DO meter)

For control and management in wastewater treatment and production processes.

Ideal for the management of aeration tanks at facilities such as wastewater treatment plants and industrial wastewater treatment facilities. Featuring a low-cost membrane cap replacement system and safe inner liquids with KCl (potassium chloride).

DO

Abteilung: Process Water
Produktionsfirma: HORIBA Advanced Techno Co., Ltd.

Low-cost sensor replacement

For sensor replacement, only the membrane cap needs to be replaced, making it more cost-effective compared to replacing the sensor chip.

Safe inner liquid

Utilizing a safe KCl solution as the inner liquids.

Product name Dissolved oxygen converter
Model HD-200
Combined sensor 5505, 5510 (A bipolar diaphragm type polarography sensor)
Combined probe DO-1100
Measurement
range
Dissolved oxygen
concentration
0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L)
Saturation degree 0% to 200% (Display range: 0% to 200%)
Temperature 0℃ to 50℃ (Display range: -10℃ to 110℃)
Display
resolution
Dissolved oxygen
concentration
0.01 mg/L
Saturation degree 1%
Temperature 0.1℃
Performance Dissolved
oxygen
concentration
Repeatability Within ±0.5% of the full scale (response for equivalent input)
Linearity Within ±0.5% of the full scale (response for equivalent input)
Temperature Repeatability ±0.5℃(response for equivalent input)
Linearity ±0.5℃(response for equivalent input)
Transmission
output
Number of output 2 (The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 900Ω
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Output 1 Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range.
Output 2 Free setting within a range between -20℃ and 130℃
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact output Number of output 3
Output type No-voltage contact output
Contact type Relay contact; SPDT (1c)
Output capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact
function
R1 and R2 Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding of transmission output, and cleaning output (The contact is closed during alarm operation, opened normally and while the power is down).
FAIL Error alarm (Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting range Selectable from dissolved oxygen concentration, and temperature
  • Setting range: Within the measurement range
  • Delay time: 0 to 600 seconds
Control setting range
  • Setting range: 0 mg/L to 20 mg/L
  • Control width: 0.02 mg/L to 200 mg/L(±0.01 to ±0.50 mg/L)
Cleaning
output
Number of output points 1
Output type Contact output with voltage (output of connected power supply voltage)
Contact type Relay contact; SPST (1a)
Output contact capacity 250 V AC 0.5 A
External instrument Solenoid valve for cleaning control
Settings Cleaning period 0.1 hours to 168.0 hours
Cleaning time 2 seconds to 600 seconds
Hold time 2 seconds to 600 seconds
Timer accuracy Within 2 minutes per month
Description of cleaning
operation
One of the following operations.
  • Operation of internal timer only
  • Operation of internal timer and external input command
  • Enable the internal timer only when external input is active.
  • The cleaning trigger signal (The external trigger signal must be longer than 2 seconds)
Contact input Number of input 1
Contact type No-voltage “a” contact of open collector
Conditions ON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact function External input for cleaning or transmission holding if cleaner is not attached.
Transmission
capability
Communication type RS-485
Signal type 2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
(The temperature sensor is built into the probe DO-1000.)
Temperature compensation
range
0℃ to 50℃
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Calibration method Span calibration in the air or DO saturated aquaous solution
Number of calibration points Atmospheric calibration: 1 point
(Zero electric calibration is carried out automatically)
Saturation liquid calibration : 2 points (zero calibration is omitted)
Additional capabilities Salinity compensation (0.0% to 5.0%)
Automatic detection of calibration failure (Zero error, Sensitivity error)
Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity)
Self-check Calibration error Zero calibration error, sensitivity error, outside of temperature calibration range
Electrode diagnostic error Sensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor
meter error CPU error, CPU error, ADC error, memory error
Operating temperature range -20℃ to 55℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature -25℃ to 65℃
Power supply Rated power supply voltage 100 V to 240 V AC ±10% 50/60 Hz
Power consumption 15 VA (max)
Others With power switch for maintenance use
Compatible
standards
CE marking EMC  :EN61326-1
      Class A, Industrial electromagnetic environment
Safety  :EN61010-1
RoHS :EN50581
      9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Mounted on 50A pole or wall
Protection code IP65 IEC60529, JIS C0920
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions   180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass   Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg

*1:   The standard for effect on the reading by the electromagnetic field of the radiated radio frequency and by the conducted interference is within the measured dissolved oxygen value ±0.4 mg/L.

*2:   When the sensor cable, the transmission cable, or the contact input cable is extended to 30 m or longer, the surge test specified in the EMC directive for CE marking is not applied.

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.