3D fluorescence spectroscopy records fluorescence intensity across multiple excitation and emission wavelengths to produce an excitation–emission matrix (EEM). This spectral fingerprint helps characterize fluorescent samples and compare complex mixtures. Absorbance measurements support inner-filter effect correction, while chemometric methods help interpret overlapping signals.
Abstract
How can you distinguish fluorescent components when their signals overlap? A single excitation or emission spectrum may reveal only part of the information available in a complex sample.
In Part 2 of HORIBA’s fluorescence education series, Salauat Kiraev introduces 3D fluorescence spectroscopy and excitation–emission matrices (EEMs). Discover how recording emission spectra at multiple excitation wavelengths creates a molecular fingerprint that supports sample comparison and mixture characterization.
The webinar explains the main components of a spectrofluorometer and compares photomultiplier tube and CCD detection for EEM acquisition. It then examines primary and secondary inner-filter effects, showing how absorption and reabsorption can distort fluorescence intensity and spectral shape.
Learn how A-TEEM spectroscopy combines absorbance, transmittance and fluorescence EEM measurements to support inner-filter effect correction. Explore how parallel factor analysis (PARAFAC) and principal component analysis (PCA) help interpret complex datasets.
Examples cover water characterization, petroleum-related contamination, wine oxidation, cell culture media classification, insulin characterization and skin fluorescence before and after sunscreen application.
Watch the webinar to understand how corrected fluorescence fingerprints and multivariate analysis can provide more information from complex samples.
What you will learn
Who should watch?
Researchers, analytical scientists, students and laboratory teams working with fluorescent mixtures, particularly in environmental analysis, food and beverages, biotechnology, pharmaceutical research and cosmetics.
Language: English | Duration: Approximately 40 minutes
Series: What is Fluorescence?
An excitation–emission matrix, or EEM, records fluorescence intensity for combinations of excitation and emission wavelengths. It is commonly displayed as a contour map or three-dimensional surface, showing how the sample responds across both wavelength ranges.
A conventional emission spectrum uses one excitation wavelength. An EEM combines emission spectra collected at multiple excitation wavelengths. This matrix provides a broader spectral fingerprint and can reveal differences that are less apparent in a single spectrum. The measured EEMs contain both emission and excitation spectral information.
A-TEEM combines absorbance, transmittance and fluorescence excitation–emission matrix measurements. The absorbance data support correction of inner-filter effects, helping obtain fluorescence fingerprints that are more suitable for sample comparison and chemometric analysis.
The primary inner-filter effect occurs when the sample attenuates the excitation light before it reaches the observed measurement region. The secondary effect occurs when emitted fluorescence is reabsorbed before reaching the detector. Both can reduce measured intensity and distort spectral profiles.
Without correction, concentration-dependent optical effects can appear as changes in sample composition. The webinar illustrates this with gallic acid: apparently different EEM profiles become more similar after correction. Appropriate correction helps separate optical distortion from genuine sample differences.
In suitable conditions, correction can extend the useful concentration range and reduce the need for dilution. It does not remove every measurement limitation. Highly absorbing, scattering or otherwise complex samples still require an appropriate measurement strategy and validation.
In a scanning PMT configuration, emission wavelengths are measured sequentially. A CCD can record multiple emission wavelengths during one exposure, potentially shortening EEM acquisition. Actual performance depends on the spectral range, instrument configuration and required signal quality.
Parallel factor (PARAFAC) analysis decomposes a collection of EEMs into modelled components with characteristic excitation and emission profiles and sample-dependent contributions. These components can help interpret overlapping fluorescence. Assigning a component to a specific compound requires supporting evidence.
Principal component analysis (PCA) summarizes the main variation in a dataset. It can reveal sample groupings, differences and potential outliers. In the webinar, PCA supports the comparison and classification of cell culture media.
Examples include natural organic matter in water, petroleum-related fluorescence, wine oxidation and preservation, cell culture media, insulin samples, and changes in skin fluorescence after sunscreen application.
Salauat Kiraev is a Fluorescence Product Specialist at HORIBA France. In this second webinar of the fluorescence education series, he explains EEM acquisition, inner-filter effects and multivariate analysis through practical examples involving complex fluorescent samples.
Read more about Salauat Kiraev 〉