EL300

Multi-Channel Water Controller

The EL300 is a universal water controller based on a modular concept.

The EL300 can adapt to many different probes and configurations, mono or multi-channel. It allows the connection of one or several probes for each parameter among pH, ORP, dissolved oxygen, conductivity, chlorine, total suspended solid (TSS), total dissolved solid (TDS), temperature and turbidity in the limit of 16 channels. A very user-friendly interface can display all the values as well as graphs of the recorded values with periods from 10 minutes to one month. A USB port allows to transfer the recorded measurements that may be imported to Excel for treatments or graphs. A new web-based interface allows the control and the troubleshooting at distance using an internet browser on a computer, tablet or iPhone.

Quenching fluorescence based oxygen probe

  • The dissolved oxygen probe is based on the fluorescence method for a lower maintenance and higher stability.
  • At the opposite of galvanic and polarographic probes, the fluorescence based probes requires no electrolyte refill, no membrane change and no routine calibration. No flow is needed because there is no oxygen consumption.
  • They also perform very well in harsh environments that normally destroy other conventional sensors.

 

Robust industrial probes

All the probes are specially designed for harsh environments with high level of suspended solid.

ORP/pH probe

Dissolved oxygen probe

pH differential probe

Total suspended solid probe

Turbidity probe

Conductivity probe

 

User-friendly interface

The colour touch screen and intuitive interface available in 8 different languages (Chinese, English, French, German, Italian, Portuguese, Spanish, Turkish) makes very easy to test or configure the analyser. Many test functions allows to test and troubleshoot each element of the analysers (light signal, pumps, solenoid valves, etc...) to setup quickly a maintenance diagnostic. The complete configuration can be saved on a USB key and reload if necessary.

 

Communication

Additional parameters are available like status code, error code, calibration values and pumps run time. Basic 4-20 mA output modules can be plugged on the main board for each measuring channel, in the limit of 12 modules. A USB port enables to download on any USB key the last 5000 recorded measurements as well as a diagnostic file containing the configuration and useful information for remote troubleshooting. The new web interface makes possible to drive remotely the analyser from any computer, tablet or i-phone with a web browser. For this, an optional Wi-Fi or Ethernet module is added inside the analyser to connect it to an existing network with an internet gateway. The recorded measurements file can be imported to Excel for graphs or other treatments. The software of the analyser can be upgraded by connecting a USB key.

Low maintenance and high reliability

The EL300 delivers a periodic 12V output to drive an air compressor to clean the probes equipped with air cleaning. The IP65 enclosure with an acid resistant protection film on the screen assume a efficient long term protection of the analyser.

 

Modular concept

The EL300 main electronics board includes 12 sockets that can receive input or output modules.

Input modules concern mainly pH and conductivity, while output modules can be 4-20 mA output or relays for high or low alarms. A RS485 port directly connects to a serie of RS485 probes for dissolved oxygen, turbidity (high range and low range) or pH with long cable distance. Several probes can be connected in parallel on the same RS485 port. A scan function in the software automatically recognises the network of probes connected.


Multi-channel water controller, EL300 - Parameter specification

ParameterStandard Range Other ranges on request Typical RepeatabilityAccuracy
(On Standard Solution)
pH0-14+/- 0.01 pH+/- 2%
ORP+/-2000 mV+/- 1 mV+/- 2%
Dissolved oxygen0-25 mg/l O2+/- 0.02 mg/l O2+/- 2%
Conductivity0-20 μS (K=0.01)
0-200 μS (K=0.1)
0-2000 μS (K=1)
0-20 mS (K=10)
+/- 0.01 μS
+/- 0.1 μS
+/- 1 μS
+/- 0.01 mS
+/- 2%
+/- 2%
+/- 2%
+/- 2%
Total Residual Chlorine0-5 mg/l+/- 0.01 mg/l+/- 2%
pH0-14+/- 0.01 pH+/- 2%
Turbidity
(TSS by correlation)
0-4 NTU
0-40 NTU
0-400 NTU
 +/- 2%
+/- 2%
+/- 2%
TSS (Total Suspended Solid)0-1500 mg/l TSS
0-30000 mg/l TSS
+/- 1% of reading or +/- 2 mg/l TSS
+/- 1% of reading or +/- 2 mg/l TSS
+/- 2%
+/- 2%
TDS (Total Dissolved Solid)0 – 20 mg/l
0 – 200 mg/l
0 – 2000 mg/l
0 – 20 g/l
+/- 0.01 mg/l
+/- 0.1 mg/l
+/- 1 mg/l
+/- 0.01 g/l
+/- 2%
+/- 2%
+/- 2%
+/- 2%
Temperature0-80°C+/- 0.1 °C+/- 2%
+/- 2%

 

Multi-channel water controller, EL300 - General specification

Sample temperature0 - 60 °C
Measuring time< 5 sec
Measurement intervalContinuous or periodic, 1 min to 720 min
Memory5000 lines of measurements (up to 16 channels) with date and time
Power supply90- 264 VAC 50/60 Hz 40 VA - 12V DC 3A maxi
ScreenColour TFT LCD 320x240 pixels with LED backlight
CommunicationRS232, MODBUS or HTTP/Web interface (Window 7 with IE9, Android with Opera, Apple i-phone with Safari)
RS485 for probes (DO, TSS)
USB
Wi-Fi (IEEE802.11B) optional
Ethernet (IEEE802.3) optional
CertificationsCE, EN 61010-1, EN 61326
EnclosureStainless steel with epoxy coating, IP65, wall mounting brackets
Dimensions300 x 200 x 110 mm
Weight6 kg approx.

 

Multi-channel water controller, EL300 - Accessories

Input modules
Product code: MI4-20

4-20 mA input module
Isolated 4-20 mA output
Impedance: 100 Ohm
 

Input modules
Product code: MIL

Double logical inputs module
Input no 1 : external pulse command for measurement
Input no 2 : measurements inhibition
Isolated 0 – 48 V DC inputs
Impedance: >10 Kohm
 

Output modules
Product code: MO4-20

4-20 mA output module
Isolated 4-20 mA output
Active output, Max load 500 Ohm
 

Output modules
Product code: MRELAY

Relay module
Contact rating: 2A/220V
Maximum 6 relays modules allowed

Požadavek na informaci

Máte nějaké dotazy nebo požadavky? Pomocí tohoto formuláře kontaktujte naše specialisty.

* Tato pole jsou povinná.

Corporate