| General specification | Name of equipment | Automatic COD monitor |
| Model | CODA-500 |
| Object | COD concentration in water |
| Dimensions | 600 (W) mm × 510 (D) mm × 1600 (H) mm*1 |
| Mass | Approx. 150 kg |
| Power source | 100 V to 240 V AC ±10%, 50 Hz/60 Hz |
| Power consumption | - 100 V to 120 V AC: Approx. 250 VA
- 120 V to 240 V AC: Approx. 350 VA
|
| Installation conditions | - Indoor installation type
- Transient overvoltage of main power source: Overvoltage category II (EN61010-1), pollution level 2
|
| Performance | Measurement range (Actual upper limit of measurement is half-scale of the range.) | - 0 mg/L to 20 mg/L
- 0 mg/L to 30/40/50/100/200/500/1000/2000 mg/L (1-dilution type)
|
Reproducibility (with standard solution for glucose) | 20 mg/L: Within ±1% of full-scale value 30 mg/L to 500 mg/L: Within ±2% of full-scale value 1000 mg/L to 2000 mg/L: Within ±5% of full-scale value Optional specification*2: Within ±5% of full-scale value |
| Stability | Zero drift (for 24 h) | 20 mg/L: Within ±3% of full-scale value 30 mg/L to 500 mg/L: Within ±4% of full-scale value 1000 mg/L to 2000 mg/L: Within ±5% of full-scale value Optional specification*2: Within ±5% of full-scale value |
Span drift (for 24 h, with standard solution for glucose) | 20 mg/L: Within ±3% of full-scale value 30 mg/L to 500 mg/L: Within ±4% of full-scale value 1000 mg/L to 2000 mg/L: Within ±5% of full-scale value Optional specification*2: Within ±5% of full-scale value |
Measuring principle*3 | CODA-500-A (acidic method) | Acid potassium permanganate method at 100℃ (based on JIS K 0806) |
CODA-500-B (alkaline method) | Alkaline potassium permanganate method at 100℃ |
CODA-500-C (acidic method) | Acid potassium permanganate method at 100℃ (based on JIS K 0806) With reagent cleaning function |
| Number of measurement points | Standard: 1 Optional: 2 |
| Measuring range | Standard: 1 range Optional: 2 ranges |
| Heating method | Direct heating |
| End point detection | Potentiometric titration at constant current |
| Titration method | Micro syringe titration |
| Measurement interval | 60 minutes |
| Reagent cleaning | CODA-500-A, CODA-500-B: Not available CODA-500-C: Available |
Method without adding a silver nitrate solution*4 | Available (by selecting this method when the chloride ion concentration is low) |
Measuring conditions | Ambient temperature | 2℃ to 40℃ |
| Ambient humidity | Relative humidity: 85% max. (without condensation) |
Power supply voltage fluctuations | 100 V to 240 V AC ±10% |
Sample water conditions | Temperature | 2℃ to 40℃ (without freezing) |
| Flow rate | When the overflow tank (optional) is used 0.5 L/min to 5 L/min (depends on the overflow tank type) |
Chloride ion concentration (for acidic method) | CODA-500-A: Up to 1 time the full-scale value CODA-500-C: Up to 100 times the full-scale value (maximun)*6 (The alkaline method is not influenced by chloride ion concentration. For more than 100 times the full-scale value, select the alkaline method.) |
| Sampling point | Piping length from main unit: Within 3 m |
Blank water conditions | Supply method | Standard: Waterworks water Optional: Supply from deionized water tank |
| Water quality*5 | Waterworks water without COD components (hardness: 100 mg/L max.) |
Water supply pressure | 100 kPa to 500 kPa (at the connection of the instrument inlet) |
| Consumption | 20 mL to 420 mL per measurement run (depends on the measurement range and cleaning function setting). |
| Installation conditions | - Well ventilated indoor location without exposure to direct sunlight
- Flat and stable location with minimized vibrations and shocks
- Atmosphere free from dust, mist, corrosive gas, etc.
|
Input/output specification | Display | Color LCD with touch panel |
Analog output | Number of points | Standard: 3 Optional: 6 |
| Kinds | Standard 4 mA to 20 mA DC, 0 mA to 16 mA DC (4 mA to 20 mA DC is set at the factory. 0 mA to 16 mA DC is selectable on the screen.) Optional: 0 mA to 16 mA DC, 0 V to 1 V DC, and 1 V to 5 V DC |
| Description | COD concentration, time COD load, and time flow rate |
| Output impedance | 900Ωmax. |
Contact output | Number of points | Standard: 14 Optional: 21 |
| Format | No-voltage contact output |
| Kinds | Insulated output |
| Output capacity | 250 V AC, 3 A 30 V DC, 3 A (Only 30 V DC, 3 A are available for "Maintenance" contact.) |
| Status output | Meas., Cal., Standby, Cleaning, Blank Meas., Synchro. Idle 1, Synchro. Idle 2, Maintenance, Power, etc. |
| Warning output | Limit warnings (COD Hi limit, Flow limit, and COD Hi Load), COD H.Hi limit, Sample Lack, Total Alarm 1 to 6, various gauge errors, etc. (optional: Lack Reagent) |
Analog input | Number of points | Standard: 1 Optional: 2 |
| Kinds | Standard: 4 mA to 20 mA DC input Optional: 1 V to 5 V DC |
| Description | Flow signal (full-scale value may be specified freely) |
Contact input | Number of points | Standard: 9 Optional: 17 |
| Format | No-voltage a-contact input (open collector is available) |
| Kinds | Insulated type input: common to (-) side |
| ON resistance | 100Ω max. |
| Open voltage | 26 V DC max. |
Short-circuit current | 13 mA DC max. |
| Functions | Meas. Start, Cal. Start, Cleaning Start, Blank meas. Start, Modify Date, Samp. Lack, Line Select, Flow Mainte, Flow Err., Flow Power OFF, and Flow No Drain |
| Load calculation capability | COD load calculation capability |
| Storage capability | Data retention for 1 year (measurement values) External USB storage is available. |
| Communication capability | Standard: RS-485 Optional: RS-232C |
| Communication protocol | Modbus |
| Printer | Provided (58 mm); with automatic take-up device |
Reagent consumption *7 | Potassium permanganate Reagent A | 2 mL per measurement run |
Sodium oxalate Reagent B | 1 mL per measurement run*8 |
Sulfuric acid solution Reagent C | 1 mL per measurement run |
Silver nitrate solution Reagent D | 0.5 mL per measurement run (used for acidic method) |
Sodium thiosulfate solution Reagent E | 2 mL per measurement run (used for reagent cleaning function) |
Sodium hydroxide solution Reagent F | 1 mL per measurement run (used for alkaline method) |
| Reagent replacement interval | 1 month or more. |
| Tap water | 20 mL to 420 mL per measurement run (depends on the measurement range) |