Turbidity Sensor

Our turbidity sensor is based on nephelometric method (ISO 7027 – EN 27027). A light beam from a laser diode is sent to the sample through an optical lens. The 90 degree scattered light by suspended particles is collected by a photo detector and converted to an electric signal proportional to the turbidity of the sample.
The sensor uses a 650 nm red laser light or on request a 860 nm infrared laser light so that the measuring is not affected by the color of the sample (ISO7027). It is suitable for measurement of turbidity in drinking water and aquaculture as well as for the outlet of municipal or industrial water treatment plants.
A integrated bubble trap avoid the effect of bubbles on the measurement.
The sensor can be easily opened for calibration on a formazine solution. A valve at the bottom allows to empty the cylinder during calibration operations or for a manual cleaning.
The turbidity sensor must be connected to the EL200 controller for display, calibration and measurement transmission though RS232/RS485 under Modbus protocol or by analog 4-20 mA output.

Measurement methodNephelometric
Measurement range0 - 100 NTU (other ranges on request)
Detection limit0.015 NTU
Repeatability+/- 0.01 NTU
Accuracy+/- 2% or +/- 0.015 NTU whichever is greater up to 20 NTU, +/- 5% above 20 NTU
Detector90ᵒ scattered light detector
Light sourcevisible laser diode (or infrared laser diode on request), lifetime > 5 years
Bubble Trapintegrated
Capacity250 ml
Inlet/outlet fittingsfor 9.6 mm external diameter plastic tubing
Ambient temperature range-5 to 60 ᵒC
Working pressure6 Bar maximum at 25 ᵒC
Body materialPOM
Wet materialPOM, PP, FKM, quartz
Sensor cable Length3 meters (to EL200 controller)
Associated controllerEL200 (to order separately)
Mountingon wall by 4 screws
Sensor warranty1 Year

 

Reference part 

TURB200Nephelometric turbidity sensor low range
Range: 0 – 100 NTU
3 meters of cable

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.