General specifications | Equipment name | SILICA ANALYZER |
|---|
| Model | SLIA-5000 |
| Dimensions | 600 (W) mm × 750 (H) mm × 1600 (D) mm |
| Mass | Approximately 130 kg |
| Power supply | 100 V AC ±10%, 50/60 Hz (Option) 110 V AC/115 V AC/120 V AC/220 V AC/230 V AC/240 V AC/ 240 V AC±10%, 50/60 Hz |
Power consumption | Approximately 500 VA |
| Performance | Measurement range | - 0 μg/L to 10 μg/L/0 μg/L to 100 μg/L
- 0 μg/L to 50 μg/L/0 μg/L to 500 μg/L
- 0 μg/L to 100 μg/L/0 μg/L to 1000 μg/L
- 0 μg/L to 200 μg/L/0 μg/L to 2000 μg/L
- 0 mg/L to 0.50 mg/L/0 mg/L to 5.00 mg/L
(Either is specified when receiving the order)
|
|---|
Resolution (Unit setting: μg/L) | 0.1 μg/L | Data<100 μg/L |
| 1 μg/L | 100 μg/L≦Data≦5000 μg/L |
Resolution (Unit setting: mg/L) | 0.0001 mg/L | Data<0.1 mg/L |
| 0.001 mg/L | 0.1 mg/L≦Data≦5.0 mg/L |
| Repeatability | ±2% of the full scale |
Measurement principle | Molybdenum blue method |
Number of measurement points | 1 point to 4 points (specified when receiving the order) |
| Measurement time | 5 minutes |
| Measurement cycle | Specified as desired between 5 minutes and 999 minutes |
| Calibration | Zero | Each measurement |
| Span | Equalizing filter method |
Reagent replacement cycle | Once a week (reagent storage: 30℃ or below) |
Installation conditions | Ambient temperature | 0℃ to 45℃ |
|---|
| Ambient humidity | Relative humidity 85% or less (no condensation) |
Ambient environment | - Indoor installation type
- Transient overvoltage of main power supply: Overvoltage Category II
(EN61010-1), Pollution Degree 2 - Flat surface that is not subject to vibration and external shock
- No dust, mist, or corrosive gas is present in the atmosphere
- Under atmospheric pressure
- No direct sunlight
- Well-ventilated area
|
Measurement conditions | Sample conditions | Temperature | 2℃ to 40℃ (36°F to 104°F) |
|---|
| Flow rate | 0.5 L/min to 1.0 L/min |
| Pressure | 0.02 MPa to 0.20 MPa |
| pH value | 11.0 or less |
Coexisting phosphate ion concentration | 20 mg/L or less |
However, no phosphate ions are permissible for the 0 μg/L to 10 μg/L/ 0 μg/L to 100 μg/L range (ultrapure water specification) |
External input/output | Analog output | Concentration output |
|---|
| 4 mA DC to 20 mA DC or 0 mA DC to 16 mA DC, selection setting in the operation section |
| Marker/hold output, selection setting in the operation section |
| Maximum load of 500 Ω or less, input/output isolation type, nonisolated between channels |
| Option: 0 V DC to 1 V DC, 1 V DC to 5 V DC |
| Contact output | Standard | a-contact output (b-contact can be specified when receiving the order) |
| Contact rating | 250 V 3 A AC, 30 V 3 A DC |
Measurement line shared | Meas., Standby, Maintenance, External control, Sync. Signal, Equipment Err., Power, etc. |
Measurement line individual | Line, Sample Lack, Abnormal concentration, etc. |
| Contact input | No-voltage a-contact input (open collector is available) |
| Isolated input: Common to the (-) side |
| ON resistance | 100 Ω (max.) |
| Open-circuit voltage | 26 V DC (max.) |
| Short-circuit current | 13 mA DC (max.) |
Measurement line shared | Meas. Start, Cal. Start |
Measurement line individual | MEAS. LINE SET, MEAS. RANGE SET |
Equipment function | Display | Color touch panel type liquid crystal |
|---|
| Memory function | Measurement value | Memory for over 1 year |
| Alarm history | 500 events |
| Calibration history | 100 events |
| Operation history | 100 events |
| USB external memory device can be used |
| Option | Filter for sample line (For Line 1 to Line 4) | By installing about 5 μm cartridge filter in the upstream of the overflow tank, it is possible to prevent iron rust, etc. from entering the equipment. - Up to 2 filters can be built into the product.
- If the optional filter is selected for the product of 3 or 4 measurement
points, the filter is not built into the product but is delivered individually as a single filter. - If a step-down transformer is installed (power supply voltage is other
than 100 V AC), a filter cannot be built into the product.
|
|---|
Flowmeter for sample line (For Line 1 to Line 4) | By adding this, you can accurately measure the sample supply flow rate. |