Product name | Laser scanning type surface oil detector | |||
---|---|---|---|---|
Model number | LO-300 | |||
Subject of detection | Oil on water surface or liquid on floor | |||
Detection method | Surface scanning laser/reflected light sensor | |||
Detection distance | 0.3 m to 4 m (Piezometric surface Max.6 m) | |||
Laser emitter | Semiconductor laser, wavelength (λ) 650 nm, within 1 mW output (Class 2) | |||
Modulation laser freq | About 50 kHz | |||
Scanning Mechanical | Oscillation mirror | |||
Leak detection | Leak detection alert | When the state to exceed a set level of “Oil WARNING” or “OIL ALARM” passes longer than “CONT.TIME SET.”, it judges alert. | ||
System Error | Malfunction | Engages when Temp., laser emission, laser scanning, motor rotation or power fails. | ||
Abnormal light sensing | Engages when Upper light reception level, Lower light reception level, floating debris produces no reflection or abnormally strong reflection. | |||
Contact output | Number of outputs | 2 | ||
Output type | No-voltage contact output | |||
Contact type | Relay contact; SPDT(1c) | |||
Output capacity | 250 V AC 4 A,30 V DC 4 A | |||
Output 1 | AL (Oil Alarm) | When “OIL ALARM” is detected, it operates. | ||
Output 2 | ER (System Error) | When “System Error” is detected, it operates. | ||
Transmission output | Number of outputs | 1 | ||
Output type | 4 mA to 20 mA DC (Isolator) | |||
Load resistance | Max. 550 Ω | |||
Setting and Indication | (Indication) | Status of water surface | NORMAL RUN | |
OIL WARNING | ||||
OIL ALARM | ||||
HIGH LEVEL | ||||
LOW LEVEL | ||||
SYSTEM ERROR | ||||
Light reception level | 4 mA to 20 mA is transformed converted into Level bar of 17 Step. | |||
Surface Oil area | 0% to 100% is transformed converted into Level bar of 10 Step. | |||
LOG | Max. 100 points Alarm history | |||
(Setting) | OIL LEVEL | Adjustment: 4 mA to 20 mA (1 mA carving) Indication: red △ marker under LEVEL bar | ||
AREA LEVEL | Adjustment: 0% to 90% (10% carving) Indication: △ marker under AREA bar Oil warning: Yellow △ Oil alarm: Red △ | |||
CONT.TIME | Adjustment: 0 seconds to 999 seconds (1 second carving) | |||
RELAY OUT (AL) | Adjustment: MODE: Select of Oil warning or Oil Alarm TIME : 0 seconds to 99 seconds, ∞(infinity) | |||
LEVEL ERR. | Adjustment: “ON” or “OFF” | |||
MULTI-TRANS | Adjustment: “ON” or “OFF” | |||
OVER LEVEL | Adjustment: 4 mA to 24 mA | |||
DATE | Adjustment: Year/Month/Day/Hour/Minute | |||
CONTRAST | Adjustment: CONTRAST LEVEL of LCD | |||
COLOR | Adjustment: COLOR of LCD | |||
FACTORY | Adjustment: at the time of shipment from the factory | |||
Laser replacement Contact output time setting | Adjustment: 0:00 to 23:59 (unit 1 minute) | |||
Power supply | Power range | 100 V AC ±10% 50/60 Hz | ||
Power consumption | 100 W less than | |||
Other | With time-lag fuse (125 V 2 A) | |||
Operating temperature range | −10℃ to +50℃ (without freeze) | |||
Operating humidity range | Relative humidity: 5% to 95% (without condensation) | |||
Compatible standards | CE marking | EMC Directive: EN61326-1 | ||
Low-voltage Directive: EN61010-1, EN60825-1 | ||||
Structure | Installation | Outdoor installation type | ||
Protection code | IP66 corresponding | |||
Material | Detection part | AL (Chromate processing after urethane painting) | ||
Setting and Indication part | AL (Alumite processing after urethane painting) | |||
Paint color | U35-85B (Edition U of JPMA) | |||
External dimensions | φ185 mm × H547 mm | |||
Mass | Main body: Approx. 11 kg; Cables: Approx. 1.4 kg | |||
Installation condition | Installation inclination: less ±3° It avoids direct sunlight and it sets it up. Frozen provisions of the surface of the water of detection are necessary. The rain avoiding provisions are necessary. | |||
Attached cable | Power cable | 1 cable (Length:5 m NA6UCL AWG16 × 3 cores (outer diameter φ10.4)) | ||
Signal cable | 1 cable (Length:5 m NA3VCSB AWG21 (0.5 sq) × 8 cores with shield (outer diameter φ8.5)) |