Field-installation Type pH Meter (Two-Wire Transmission)
HP-300 connects pH sensor and measures pHand temperature in the sample water. Suitable for monitoring and control of various production processes and wastewater treatment, etc. Many types of electrodes are available to match the sample properties.
pH: 0 - 14
Temperature: 0 - 100 ℃
Product name | pH meter | |||
---|---|---|---|---|
Model | HP-300 | |||
Combined sensor | Glass electrode | |||
Measurable range | pH | pH0 to pH14 (display range: pH−1 to pH15) | ||
Temperature | 0℃ to 100℃ Display range when automatic detection capability for each temperature sensor type is used: −10℃ to 110℃ Display range when temperature sensor type is manually set: −20℃ to 130℃ | |||
Display resolution | pH | 0.01 pH | ||
Temperature | 0.1℃ | |||
Performance | pH | Repeatability | Within ±0.03 pH (response for equivalent input) | |
Linearity | Within ±0.03 pH (response for equivalent input) | |||
Temperature | Repeatability | ±0.3℃ (response for equivalent input) | ||
Linearity | ±0.3℃ (response for equivalent input) | |||
Transmission output | Output type | 4 mA to 20 mA DC: input/output isolated type | ||
Load resistance | Maximum: 600Ω Case of 24 V DC power supply (*Note 1) | |||
Linearity | Within ±0.08 mA (output only) | |||
Repeatability | Within ±0.08 mA (output only) | |||
Output range | pH: Selection from preset ranges or free range input within measuring range. | |||
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA | |||
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. | |||
Contact input | Number of input points | 1 | ||
Contact type | No-voltage "a" contact for open collector | |||
Conditions | ON resistance: 40 Ω Open voltage: 1.2 V Short-circuit current: 21 mA DC max. | |||
Contact function | External input for transmission holding. | |||
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) | ||
Positive relation resistor with temperature :500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃) | ||||
Element selection method | Automatic temperature sensor type detection or manual selection (No temperature element is selectable) | |||
Temperature compensation range | 0℃ to 100℃ | |||
Temperature calibration | 1 point calibration comparing reference thermometer | |||
Calibration | Calibration method | Auto or basic (manual) calibration | ||
Number of calibration points | Selectable from 1, 2, and 3 points | |||
Kinds of standard solutions | pH 2, 4, 7, 9, 10 | |||
Any pH standard solution (with difference of 2 pH or more) for basic calibration | ||||
Additional capabilities | Automatic detection of kind of standard solution | |||
Automatic detection of electric potential stability | ||||
Automatic detection of calibration failure (asymmetry potential, sensitivity, or response time) | ||||
Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration) | ||||
Self-check | Calibration error | Asymmetry potential error, sensitivity error, and response time error | ||
Temperature calibration error | ||||
Standard solution detection error | ||||
Electrode diagnostic error | Cracking of glass response membrane | |||
Reference electrode impedance error (only applicable for differential circuit mode with a liquid ground electrode) | ||||
Temperature sensor short-circuit, temperature sensor disconnection, and deviation from temperature measurement range | ||||
Meter error | CPU, ADC, and EEPROM errors | |||
Operating temperature range | −20℃ to 60℃ (without freeze) | |||
Operating humidity range | Relative humidity: 5% to 90% (without condensation) | |||
Storage temperature | −25℃ to 65℃ | |||
Power supply | Rated voltage | 24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 1) | ||
Power consumption | 0.6 W max. | |||
Compatible standards | CE marking | Compatible standards | EMC directive | EN61326-1 |
RoHS Directive | EN50581 | |||
Immunity | Industrial electromagnetic environment | |||
Electostatic discharge | IEC61000-4-2 | |||
Electromagnetic field of radiated radio frequency | IEC61000-4-3 (*Note 2) | |||
Electric fast transient/burst | IEC61000-4-4 (*Note 2) | |||
Surge | IEC61000-4-5 (*Note 3) | |||
Conducted interference induced by radio frequency | IEC61000-4-6 (*Note 2) | |||
Emission | Class A | |||
Radiated disturbance | CICPR 11 CLASS A | |||
FCC rules | Part15 CLASS A | |||
Structure | Installation | Outdoor installation type | ||
Installation method | Mount on 50 A pole or wall | |||
Protection code | IP65 | IEC60529, JIS C0920 | ||
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) | |||
Material of fittings | SUS304 | |||
Material of hood | SUS304 stainless steel (coated with epoxy-denatured melamine resin) | |||
Material of window | Polycarbonate | |||
Display element | Reflective monochrome LCD | |||
External dimensions | 180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets) | |||
Mass | Main body: Approx. 2.8 kg cover and mounting brackets: Approx. 1 kg |