SPN23 Specifications
| Specification | SPCS-ONE-DL-23 | SPCS-ONE-MT-23 | SPCS-ONE-DR-23 | SPCS-ONE-MD-23 |
|---|---|---|---|---|
| Measuring principle | Laser-scattering condensation particle counting (CPC) | |||
| Measured component and range | Number concentration of solid particles: 0–10,000 up to 0–50,000 particles/cm3 after internal dilution2 | |||
| CPC counting efficiency | 50 ± 12% at 23 nm; greater than 90% at 41 nm | |||
| Sample-handling temperature | Dilute sampling: 52 °C or less Maximum permissible temperature for direct sampling: 350 °C3 | |||
| Diluted-sample temperature |
|
|
|
|
| Dilution factors |
|
|
|
|
| Particle Concentration Reduction Factor (PCRF) | 0.95 < fr (30 nm) / fr (100 nm) < 1.3 0.95 < fr (50 nm) / fr (100 nm) < 1.2 | |||
| Volatile-particle removal unit | Evaporation tube | |||
| Volatile-particle removal efficiency | Greater than 99.0% at a particle size of 30 nm and a concentration of 10,000 particles/cm3 or more | |||
| Dilution-factor accuracy | Within ±10% of the nominal dilution-factor setting4 | |||
| Operating environment | Without CLU (standard): Ambient temperature: 5–30 °C; relative humidity: 80% or less; altitude: 0–2,000 m above sea level (0–2,500 m with altitude simulation). With CLU (optional): Ambient temperature: 5–45 °C; relative humidity: 80% or less; altitude: 0–2,000 m above sea level (0–2,500 m with altitude simulation). | |||
| Power-supply voltage and frequency | 200/220/230/240 V AC (±10%, maximum 250 V), 50/60 Hz (±1.0 Hz), single-phase; specify when ordering | |||
| Power requirements | Main unit: Maximum 2.5 kVA Main unit and all optional units: Maximum 4.7 kVA | Main unit: Maximum 2.7 kVA Main unit and all optional units: Maximum 4.5 kVA | Main unit: Maximum 2.7 kVA Main unit and all optional units: Maximum 4.6 kVA | Main unit: Maximum 2.8 kVA Main unit and all optional units: Maximum 4.7 kVA |
| Dimensions5 and mass | 434 (W) × 731 (D) × 637 (H) mm Approximately 123 kg | 434 (W) × 910 (D) × 637 (H) mm Approximately 144 kg | 434 (W) × 845 (D) × 637 (H) mm Approximately 128 kg | 434 (W) × 910 (D) × 637 (H) mm Approximately 149 kg |
SPN10 Specifications
| Specification | SPCS-ONE-DL-10 | SPCS-ONE-MT-10 | SPCS-ONE-DR-10 | SPCS-ONE-MD-10 |
|---|---|---|---|---|
| Measuring principle | Laser-scattering condensation particle counting (CPC) | |||
| Measured component and range | Number concentration of solid particles: 0–50,000 particles/cm3 after internal dilution2 | |||
| CPC counting efficiency | 65 ± 15% at 10 nm; greater than 90% at 15 nm | |||
| Sample-handling temperature | Dilute sampling: 52 °C or less Maximum permissible temperature for direct sampling: 350 °C3 | |||
| Diluted-sample temperature |
|
|
|
|
| Dilution factors |
|
|
|
|
| Particle Concentration Reduction Factor (PCRF) | 0.95 < fr (15 nm) / fr (100 nm) < 2.0 0.95 < fr (30 nm) / fr (100 nm) < 1.3 0.95 < fr (50 nm) / fr (100 nm) < 1.2 | |||
| Volatile-particle removal unit | Hot catalytic stripper | |||
| Volatile-particle removal efficiency | Greater than 99.9% for C40 (count median diameter greater than 50 nm; mass density greater than 1 mg/m3) | |||
| Dilution-factor accuracy | Within ±10% of the nominal dilution-factor setting4 | |||
| Operating environment | Without CLU (standard): Ambient temperature: 5–30 °C; relative humidity: 80% or less; altitude: 0–2,000 m above sea level (0–2,500 m with altitude simulation). With CLU (optional): Ambient temperature: 5–45 °C; relative humidity: 80% or less; altitude: 0–2,000 m above sea level (0–2,500 m with altitude simulation). | |||
| Power-supply voltage and frequency | 200/220/230/240 V AC (±10%, maximum 250 V), 50/60 Hz (±1.0 Hz), single-phase; specify when ordering | |||
| Power requirements | Main unit: Maximum 2.5 kVA Main unit and all optional units: Maximum 4.7 kVA | Main unit: Maximum 2.7 kVA Main unit and all optional units: Maximum 4.5 kVA | Main unit: Maximum 2.7 kVA Main unit and all optional units: Maximum 4.6 kVA | Main unit: Maximum 2.8 kVA Main unit and all optional units: Maximum 4.7 kVA |
| Dimensions5 and mass | 434 (W) × 731 (D) × 637 (H) mm Approximately 123 kg | 434 (W) × 910 (D) × 637 (H) mm Approximately 144 kg | 434 (W) × 845 (D) × 637 (H) mm Approximately 128 kg | 434 (W) × 910 (D) × 637 (H) mm Approximately 149 kg |
1 Model abbreviations: DL = Dilution; DR = Direct; MT = with Micro Tunnel; MD = with Micro Tunnel and Direct.
2 Determine the system dilution factor so that the particle concentration after dilution remains within the measuring range.
3 The allowable gas-temperature range at the sample-probe inlet depends on the sampling conditions because the DSU diluter temperature is limited to 350 °C or less. Please contact HORIBA for details.
4 Accuracy of the dilution-factor check using span gas.
5 Dimensions and mass exclude the transfer tube, control unit and optional units.