CS-900은 차세대 습식 공정 모니터입니다. HORIBA는 수율을 높이는 화학농도 관리 장비를 제공해 공정 최적화에 기여하고 있습니다. HORIBA는 반도체 WET 공정에서 요구되는 화학적 농도 제어를 충족하기 위해 기존 센서 설계를 통해 안정적인 측정과 작업자 안전성을 유지하면서 소형 장비 설계로 니즈에 맞는 보다 높은 기능성을 실현하고 있습니다.

Contact us
Find us
Zero contamination is made possible by directly connecting a sensor externally to the PFA piping. This removes the need for plumbing construction when attaching the sensor, which reduces chemical accidents. In addition, by upgrading the optical system, the equipment has been made more compact while retaining the original specifications of CS-600F*1.
Light source and light detector have been integrated into cell unit.
Panel mount display (9.6cm x 9.6cm)
Use of LED lamp Increases lifespan (5 years warranty)
Semi-annually BG compensation cycle
Piping installation is not required when installing the product, which helps to reduce accidents such as chemical leak.
*1 CS-600 is HORIBA's optical liber type chemical concentration monitor.

Stability
Measurement Condition
Measurement Frequency: every 3 seconds
Moving Average: 16 times
Temperature: room temperature
Pipe Diameter: 3/4 inch
Initial Sample Concentration
NH3 : 0.95% H2O2 : 5.15% H2O: 93.9%
Measurement result
Sample (real liquid) measured for 1 hour shows excellent stability.



| Model name | Non-contact type chemical concentration monitor | ||||
|---|---|---|---|---|---|
| Model | CS-900 | ||||
| Measurement method | Absorption spectroscopy | ||||
| Concentration calculation | Temperature compensated multivariate analysis | ||||
| Measurement range/ Repeatability (mass%) | Model name | Component | Measurable range | Repeatability | Analog output range |
| CS-900-1P-1P25P5 | NH3 | 0.00%-1.20% | ±0.05% | 0.00-10.00 | |
| H2O2 | 0.00%-5.50% | ±0.10% | 0.00-10.00 | ||
| H2O | 93.3%-100.0% | ±1.0% | - | ||
CS-900-1P-P8P8 | NH3 | 0.20%-0.80% | ±0.03% | 0.00-1.00 | |
| H2O2 | 0.20%-0.80% | ±0.03% | 0.00-1.00 | ||
| H2O | 98.4%-99.6% | ±0.3% | - | ||
Conditions of measurement | 1) Measurement Frequency: every 3 seconds 2) Moving average : 16 times | ||||
Tube | PFA tube 1inch or 3/4inch (Do NOT use anti-static type PFA tube) | ||||
Sample condition*1 | Sample temperture : 20 to 80°C Sample temperture, Atmospheric temperture fluctuation : ±1 degree C (period of time: 1 hour) Sample pressure : Less than 0.2 MPa (G) Sample pressure fluctuation : Within 0.02 MPa | ||||
Power source | DC 24V | ||||
Power consumption | Apporox. 30W (Transient electric current at the time of start is excluded) | ||||
| Weight | Display unit : Apporox. 0.9kg Sensor unit: Apporox. 2.9kg | ||||
| Dimension (WxDxH) | Display unit : 96x135x96mm Sensor unit: 238x135x190mm | ||||
| Slanting angle | Sensor unit: Vertical direction within ±3 degree Lateral direction 0 to 90 degree | ||||
| Light source | 5years | ||||
| communication | Parallel I/O, RS-232C, Analog output | ||||
HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.