HU-200TB-W

현장 설치형 탁도계 (4-Wire Transmission, For Wide Range)

HU-200TB-W는 급수장 취수구, 침전조 등 혼탁도 변화가 큰 공정에서 혼탁도를 지속적으로 측정하기 위해 개발된 혼탁도 측정기입니다. 투과 90도 산란광 방식을 채택하고 있으며 낮은 범위(10도)에서 높은 범위(2000도)까지 혼탁도 변화가 큰 공정의 변화를 모니터링하는 데 적합합니다.

옵션으로 전기 와이퍼 클리닝을 설치할 수 있으며 주기적인 클리닝을 통해 장기간 안정적인 측정이 가능합니다.

측정범위

0 ~ 1000도(度) (표준용액 : Kaolin)
0 ~ 2000도(度) (표준용액 : Formazin)
0 ~ 100도(度) (표준 용액 : PSL)

Tough

  • Responds to harsh conditions at the site process
      Operating temperature range(Transmitter):-20 - 55℃
      Rated power supply voltage: AC 100  - 240 V ± 10 % 50/60 Hz
      Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Sensor error can be detected immediately by self-diagnosis function
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring

Converter

Product name Turbidity meter
Model HU-200TB-W Industrial-use turbidity converter
Turbidity sensor
type
SS-120-W Industrial-use turbidity detector
Measurement range Kaolin 0 to 1000 (degrees) (Display range: 0 to 1100 (degrees)) or 0 to 1000 (mg/L)
Formazin  0 to 2000 (Display range: 0 to 2200 )
PSL 0 to 100 (degrees) (Display range: 0 to 110 (degrees))
Transmission output
setting range
Kaolin Arbitrary setting within the range from 0.00 to 10.00 (degrees), 0.0 to 100.0
(degrees), 0 to 2200 (degrees)
Formazin (NTU)
PSL
Display resolution
  • 0.01 degrees (Degree range: 0 to 10 (degrees))
  • 0.1 degrees (Degree range: 0 to 100 (degrees))
  • 1 degrees (Degree range: 100 to 2200 (degrees))
    Selectable between fixed range (decimal point) and automatic range changeover
Performance Repeatability Within ±2% of the reading or ±0.04 degrees, whichever is larger (checked using the solid span bottle)
Linearity The deviation at the mid-range point of the span calibration value is within ±2% of the span calibration value or ±0.04 degrees, whichever is larger
Accuracy Within ±2% of reading
Transmission output

Number of current outputs:2

Output type 4 mA to 20 mA DC, input/output isolation type
Load resistance 900 Ω (max.)
Linearity Within ±0.08 mA (limited to the current linearity)
Repeatability Within ±0.02 mA (limited to the current repeatability)
Error output Provided with burnout function (3.8 mA or 21 mA)
Hold function Selectable between the latest value hold and the preset value hold
Contact output Number of output points: 6
Output type No-voltage output contact
R1, R2 Contact type Relay contact, SPST (1a)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, cleaning output and automatic calibration (closed at alarm operation, open under normal conditions and in a power-off state)
Description of alarm
  • Setting range: Within turbidity measurement range
  • Delay time: 0 to 600 (s)
R3 Contact type Relay contact, SPST (1a)
Contact capacity 30 V DC 1 A (resistance load)
Contact function Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, cleaning output and automatic calibration (closed at alarm operation, open under normal conditions and in a power-off state)
Description of alarm
  • Setting range: Within turbidity measurement range
  • Delay time: 0 to 600 (s)
FAIL Contact type Relay contact, SPDT (1c)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function
Error alarm (closed under normal condition, open in an error or power-off state)
Description of alarm
  • Error alarms for values outside the measurement range,
    self-checks, and calibration errors can be set
  • Delay time: 0 to 600 (s)
RANG1,
RANG2
Contact type Relay contact, SPST (1a)
Contact capacity 30 V DC 1 A (resistance load)
Contact function Status output of the transmission output range
Control output for
automatic zero
calibration (ZERO
OUT, SAMPLE
OUT)
Number of output points 2
Output type Voltage contact output (connection power supply voltage output)
Contact capacity 250 V AC 0.5 A
Contact function SAMPLE OUT: Electric ball valve control output
ZERO OUT: Solenoid valve control output
Contact input Number of input points 4
Contact type Open collector no-voltage “a” contact
Condition ON resistance: 100 Ω max.
Open-circuit voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact
function
EXT1 Switchable between automatic zero calibration command and transmission hold
EXT2 Switchable between cleaning command and transmission hold
EXT3, EXT4 Selection command from four ranges using 2-bit contacts
Communication
function
Type RS-485
Signal form Two wire type, the power source of the communication is isolated from the power of measurement circuit. (The power source of transmission and communication are not isolated.)
Calibration Calibration method Zero calibration By filtered clean water
Automatic zero
calibration
Zero calibration is performed automatically by switching to the filtered clean water (option)
Span calibration Turbidity adjustment method by coefficient input
Compatible standard substances Kaolin, Formazin (NTU), PSL
Automatic zero
calibration (option)
Zero calibration
method
The electric valve of a sample line is closed, the electromagnetic valve of a zero line is opened, measured value is automatically judged to be zero within a time, and zero calibration is carried out.
Setting Calibration
cycle
1 to 999 (h)
Stability time 1 to 60 (s)
Stability level 0.01 degrees/NTU to 0.05 degrees/NTU
Hold time 60 s to 600 s
Cleaning function
(option)
Cleaning method Electric wiper type (executes cleaning operation by
communicating with the converter)
Setting Cleaning cycle 0.1 to 168.0 (h)
Cleaning time 30 to 600 s
Hold time 60 to 600 s
Timer accuracy Difference within 2 min per month
Self-check Sensor check error Sensor error
Converter error CPU abnormality, ADC abnormality, memory abnormality
Operating
temperature range
−20℃ to 55℃(no freezing)
Operating humidity
range
Relative humidity: 5% to 90% (no condensation)
Storage temperature −25℃ to 65℃
Power supply Rated power supply
voltage
100 V to 240 V AC ±10% 50/60 Hz
Power consumption 35 VA (max.) when an automatic cleaner is connected.
Others With built-in power switch for maintenance
Compatible
standards
CE marking EMC    :EN61326-1*
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS   :EN50581
      9. Industrial monitoring and control instruments
FCC rules Part 15 Class A
Structure Installation Outdoor installation type
Installation method 50 A pole-mounted or wall-mounted
Protection class IP65
Material of case Aluminum alloy (coated with epoxy-modified melamine resin)
Material of mounting
brackets
SUS304
Material of hood SUS304 (coated with epoxy-modified melamine resin)
Material of display window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 mm (W) × 155 mm (H) × 115 mm (D) (excluding brackets)
Mass Mainframe: approx. 3.5 kg (excluding brackets), hood, bracket: approx. 1 kg

*:  The surge test specified in the EMC Directive for CE Marking shall not apply to the case when a sensor cable, transmission cable, or contact input cable is extended for 30 m or longer.

Note

Although an arrestor (400 V discharge starting voltage) is mounted for the output signal, contact input, and communication, install an optimum surge absorber element on the connection line, depending on the surrounding environment, device installation conditions, and external connection devices.

Detector

Product name Turbidity detector
Model SS-120-W
Measurement principle 90-degree transmission-scattering method
Light source Near-infrared LED, 860 nm
Detector Silicon photo diode
Detection window Inside diameter: φ30 mm, hard glass tube
Data transfer RS-485 (communication with converter)
Sample water temperature 0℃ to 45℃ (no freezing)
Sample water pressure 0 MPa to 0.3 MPa
Material of liquid contact section PVC, SUS316, FKM, silicon, hard glass, EPDM
Cable length Standard accessory cable: 5 m
Installation Screwing bore size: Rc3/4
Power supply 12 V DC supplied from HU-200TB-W converter
External dimensions 131 mm (W) × 450 mm (H) × 224 mm (D) (excluding cable)
Mass Approx. 3.5 kg (excluding cable)

 