HR-200

현장 설치형 잔류염소 측정기 (4-Wire Transmission, Water flow beads)

HR-200은 잔류염소센서(RA-20)에 접속하여 유리잔류염소 농도를 고정밀도로 측정할 수 있습니다. 상수도, 수영장 물 살균 및 담수화 시스템에 적합합니다.

3극 폴라로그래픽 방식으로 시약을 사용하지 않고 잔류염소 농도를 지속적으로 측정하며 모니터링할 수 있습니다. 비드(beads)를 이용한 물리적 연마 세정으로 장기간 안정적인 측정이 가능해졌습니다.

 

측정 범위

유리 잔류 염소 농도: 0 ~ 3 mg/L
온도: 0 ~ 50 ℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.
  • Suitable for waterworks, pool water sterilization, and desalination systems
  • Long-term stable measurement with combination of bead cleaning and electrochemical cleaning
Measuring method Polarography
Measuring range Residual Chlorine: 0 mg/L to 3 mg/L   Resolution: 0.01 mg/L
Temperature: 0 to 50˚C   Resolution: 0.1˚C
Repeatability Residual Chlorine:Within ±0.05 mg/L or less   Temperature: Within ±0.5˚C(for equivalent input)
Linearity Residual Chlorine:Within ±0.05 mg/L or less   Temperature: Within ±0.5˚C(for equivalent input)
Transmission output Two points 4 to 20 mA DC   Input/output insulation type   Maximum load resistance 900 Ω
Contact output Three points No-voltage contact output Relay contact, SPDT
Contact capability R1, R2: Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding transmission output, and cleaning output.
(opened at alarm operation, closed usually, closed at power-off)
 FAIL: Error warning (normally closed; open when an error occurs; open when the power is turned OFF)
Contact input Two points Contact type: No-voltage a contact for open collector
Contact capability: Input 1: External input for transmission holding
 Input 2: Flow switch input for interlock(Open due to decreased flow)
Communication capability RS-485   Two-wire input/output insulation type (not insulated from transmission output)
Temperature compensation range 0 to 50°C
Ambient temperature 0 to 55°C
Cleaning capability Electrochemical Cleaning between Cathode and Electrochemical cleaning electrode
Calibration method Zero calibration (Zero liquid calibration)
SPAN calibration(Compare to measurement value of DPD method, Including zero electric calibration)
Self-diagnosis function Calibration error, Temperature sensor diagnostic error, Meter error
Power source 90 V to 264 V AC, 50/60 Hz 35 VA (max.) when an automatic cleaner is connected.
Construction Outdoor installation type: IP65 Protection class
Installation method: 50 A pole-mounted or wall-mounted
Material of case: Aluminum alloy (coated with epoxy modified melamine resin)
Material of mounting brackets: SUS304
Weight Approx. 4.5 kg
Equipment protected CE marking, FCC rules