HD-300

현장설치형 용존산소측정기(2-Wire Transmission), 멤브레인 형)

HD-300은 DO(용존산소) 센서(DO-1100)를 연결하여 시료수의 용존산소량을 측정합니다.
측정 방식은 멤브레인 방식의 폴라로그래피 방식입니다.
폐수 처리 또는 양식용 수족관의 DO 모니터링에 적합합니다.

- 측정 범위(용존 산소 농도) : 0 ~ 20 mg/L
- 포화도 : 0 ~ 200%
- 온도 : 0 ~ 50℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Tough

  • Responds to harsh conditions at the site process
    Operating temperature range(Transmitter):-20 - 60℃   
    Rated voltage: DC 24 V
    Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Self-diagnosis of membrane damage and infiltration into the probe
  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring
  • Span calibration by air preceded by auto-zeroing (Electrical auto zeroing function)
  • On-site membrane replacement as well as whole sensor replacement
Product name Dissolved oxygen meter
Model HD-300
Combined sensor 5505, 5510 (A bipolar diaphragm type polarography sensor)
Combined probe DO-1100
Measurement
range
Dissolved oxygen
concentration
0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L)
Saturation degree 0% to 200% (Display range: 0% to 200%)
Temperature 0℃ to 50℃ (Display range: -10℃ to 110℃)
Display
resolution
Dissolved oxygen
concentration
0.01 mg/L
Saturation degree 1%
Temperature 0.1℃
Performance Dissolved
oxygen
concentration
Repeatability Within ±0.5% of the full scale (response for equivalent input)
Linearity Within ±0.5% of the full scale (response for equivalent input)
Temperature Repeatability ±0.5℃ (response for equivalent input)
Linearity ±0.5℃ (response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 600 Ω(when power supply voltage is 24 V DC) (*1)
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact input Number of input points 1
Contact type No-voltage a-contact
Conditions ON resistance: 40 Ω maximum.
Open voltage: 1.2 V DC
Short-circuit current: 21 mA DC
Contact function External input for transmission holding.
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
(The temperature sensor is built into the probe DO-1000.)
Temperature compensation
range
0℃ to 50℃
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Calibration method Span calibration in the air or DO saturated aquaous solution
Number of calibration points Atmospheric calibration: 1 point
(Zero electric calibration is carried out automatically)
Saturation liquid calibration: 2 points (zero calibration is omitted)
Additional capabilities Salinity compensation (0.0% to 5.0%)
Automatic detection of calibration failure (Zero error, Sensitivity error)
Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity)
Self-check Calibration error Zero calibration error, sensitivity error, outside of temperature calibration range
Electrode diagnostic error Sensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor
meter error CPU error, CPU error, ADC error, memory error
Operating temperature range -20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature -25℃ to 65℃
Power supply Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC)*1
Power consumption 0.6 W max.
Compatible
standards
CE marking EMC  :EN61326-1
     Class A, Industrial electromagnetic environment
RoHS  :EN50581
     9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Attached to 50A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 (Epoxy degenerated melamine resin painting)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 2.8 kg
Cover and mounting brackets: Approx. 1 kg

*1: The maximum load resistance that can be connected is decided depending on the power-supply voltage.

Relation between power supply voltage and load resistance

Note

An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Corporate