HE-200R

현장 설치형 전기저항계(4-Wire Transmission)

HE-200R은 비저항센서(ERF Series)를 연결하여 시료수의 비저항과 온도를 측정합니다.
UPW 저항률 측정에 적합합니다.

- 측정 범위저항 : 0.00 ~ 20.00 MΩ·cm, 0.00 ~ 200.0 kΩ·m
- 온도 : 0.00 ~ 100.00 ℃

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Tough

  • Responds to harsh conditions at the site process   
    Operating temperature range(Transmitter):-20 - 55℃   
    Rated power supply voltage:AC 100 - 240 V ±10% 50/60 Hz   
    Robust aluminum die cast (Protection code:IP65)
  • CE Marking and FCC Compliance suitable for use at overseas sites

 

Intelligence

  • Convenient function to display liquid temperature at the same time(Transmission output available)
  • Designed to be environmentally friendly (Pb-free) in response to WEEE and RoHS
  • Well-examined temperature compensation function for UPW

 

Easy maintenance

  • Easy to view big display
  • Easy operation
  • Calibration log memory
  • Screw terminals that do not fall off are used in the terminal block for wiring
Product name Resistivity meter
Model HE-200R
Combined
sensor
2-electrode method resistivity sensor (ERF-001 Series) of cell constant 0.01/cm
Measuring range Resistivity MΩ•cm 0.000 to 2.000 0.00 to 20.00 (*Note 1)
kΩ•cm 0.00 to 20.00 0.0 to 200.0 (*Note 1)
Temperature 0℃ to 100℃ (Display range: −10℃ to 110℃)
Display
resolution
Resistivity As shown in the measuring table.
Temperature 0.01℃
Performance Resistivity Repeatability Within ±0.1% of the full scale (response for equivalent input)
Linearity Within ±0.5% of the full scale (response for equivalent input)
Temperature Repeatability ±0.1℃ (response for equivalent input)
Linearity ±0.5℃ (response for equivalent input)
Transmission
output
Number of output 2 (The negative terminals for transmission outputs are internally connected at the same electric potential)
Output type 4 mA to 20 mA DC: input/output isolated type
Load resistance Maximum: 900Ω
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Output 1 Resistivity: Free setting within a measuring range.
Output 2 Temperature: Free setting within a range between −20 and 130℃
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact output Number of output 3
Output type No-voltage contact output
Contact type Relay contact; SPDT (1c)
Contact capacity 250 V AC, 3 A; 30 V DC, 3 A (resistance load)
Contact
function
R1, R2 Selectable from upper limit alarm, lower limit alarm, and currently holding of transmission output.
(The contact is closed during alarm operation, opened normally and while the power is down.)
FAIL Error alarm
(Closed in the normal state, opened in the failure state or while the power is down.)
Alarm setting range
  • Setting range: within a measuring range
  • Delay time: 0 to 600 seconds
Contact input Number of input points 1
Contact type No-voltage "a" contact for open collector
Conditions ON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact function External input for transmission holding.
Transmission
capability
Method RS-485
Signal type 2 wire system, isolated from the input circuit
Not isolated from transmission circuit
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
Compensation method
  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95℃)
  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: +/−3%/℃)
  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)
Temperature compensation
range
0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)
Calibration Resistivity Based on parameter input of coefficient for the sensor cell constant.
Temperature Comparison to the accurate thermometer
Both deviation and coefficient of RTD are taken into account for calibration.
Other function UPW standard resistivity input Select one from the next.
182.3 kΩ•m (standard), 181.8 kΩ•m, 182.4 kΩ•m,
18.23 MΩ•cm (standard), 18.18 MΩ•cm, or 18.24 MΩ•cm
Clipping function If the measured value exceeds the user setting value for clipping, the measured data will be held at the clipping value.
Self-check Sensor diagnostic error Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range
Converter error CPU, ADC, and memory errors
Operating
temperature
range
−20℃ to 55℃ (without freeze)
Operating
humidity range
Relative humidity: 5% to 90% (without condensation)
Storage
temperature
−25℃ to 65℃
Power supply Rated power supply voltage 100 V to 240 V AC ±10% 50/60 Hz
Power consumption 15 VA (max)
Other With power switch for maintenance use
Compatible
standards
CE marking EMC    :EN61326-1
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS    :EN50581
      9. Industrial monitoring and control instruments
FCC rules Part15 Class A
Structure Installation Outdoor installation type
Installation method Mount on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS304 stainless steel (coated with epoxy denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External
dimensions
180 (W) mm × 155 (H) mm × 115 (D) mm (excluding cover and mounting brackets)
Mass Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg

Note

  • Note 1  : Without temperature compensation, 0.0-100.0 MΩ•cm(0-1000 kΩ•m) can be displayed.
  • Note 2  : When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 3  : An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Corporate