HR-480

잔류 염소 측정기 (4-Wire Transmission)

산업용 잔류 염소 측정기는 음용수 및 수영장의 항균성 검사에 사용되는 지표인 유리 잔류 염소를 순차적으로 측정합니다.
결합된 센서(RS-1)는 구동부가 없는 갈바닉 방식을 사용하여 유지보수가 편리하고 장기간에 걸쳐 안정적인 측정이 가능합니다.

측정 범위

유리 잔류 염소 농도 : 0 ~ 2 mg/L

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.
  • Ideal for managing water quality, for example of drinking water or swimming pools The HR-480 takes sequential measurements of free residual chlorine, which is an indicator used in the inspection of the anti-bacterial properties of drinking water and swimming pools.
  • Uses a galvanic cell method The galvanic cell method makes the sensors more maintainable than the polarographic method, meaning that there is no need to polish the sensors, allowing them to be used semi-permanently.
  • Can be installed on the same cell as pH electrodes, ORP electrodes and temperature sensors (Option: Sampling Rack SS-2, SS-3) As well as residual chlorine sensors, pH, ORP electrodes and temperature sensors can also be installed on the same cell, making it possible to construct a low-cost system.
  • Span correction is possible when measuring fluids with low concentrations Correction work can be carried out easily on-site, with no need to prepare a separate high-concentration correction fluid.
  • Uses lead-free solder To reduce the burden on the environment, lead-free solder has been used to mount the components on the printed circuit board.
Product name Industrial Residual Chlorine Meter
Model HR-480
Measurement method Galvanic cell method
Measuring range 0 to 2 mg/L 0.01 mg/L resolution (number of decimal places selectable between 1 and 2)
Repeatability Within ±0.05 mg/L (with equivalent input)
Linearity Within ±0.05 mg/L (with equivalent input)
Transmission
output
4 mA to 20 mA DC, input/output isolated type
Maximum load resistance 900 Ω
Transmission output range: freely selectable within the measurement range.
Contact output Outputs: 2 points
Alarm contact output (R1 and R2)
  Contact type: relay contact, SPDT (1c)
  Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
  Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance.
Calibration function Zero calibration (zero adjustment omissible)
Span calibration (calibration by compariing with DPD tester)
Transmission
output hold feature
Previous value hold
Optional value hold
Continuous
Selectable from the above (However, only the previous value hold is available in the
maintenance mode.)
Self-diagnosis
function
Calibration function
Zero error
Sensitivity error
Outside of the measuring range
Converter error
Ambient
temperature
−5℃ to 45℃
Relative humidity 20% to 85% (without dew condensation)
Storage
temperature
−25℃ to 65℃
Power supply Rated voltage 100 V AC to 240 V AC, 50/60 Hz, 10 VA (max)
Structure Indoor-use panel installation type
    Panel case: ABS, Terminal: PBT
Panel: IP65 dust and water proof structure
Protective structure Panel: IP65 (IEC60529, JIS C0920)
Rear case: IP20
Terminal: IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1)
Conforming
standards
CE Marking EMC Directive: EN61326-1
Low Voltage Directive: EN61010-1
FCC Rule FCC Part15
External
dimensions
48 (W) mm × 96 (H) mm × 115 (D) mm
Case depth: approx. 105 mm
Mass Approx. 400 g
Compatible sensor RS-1
Measuring target Free residual chlorine
Water sample
requirements
pH6 to pH8 (constant)
Temperature 0℃ to 40℃ (constant)